ES51932
22,000 Counts Auto DMM
1.4 Current measurement for clampmeter
ES51932 has 2 automatic and 4 manual current measurement modes for clampmeter.
The following table summarizes the full-scale range of each mode. When ES51932
operate in the automatic mode1&2, it takes high input from IVSH/IVSL (higher
range/lower range), low input from SGND and reference voltage from VR. When
ES51932 operates in the manual mode1~4, It takes high input from ADP, low input
from SGND and reference voltage from VA+ and VA-.
Mode
Automatic1
Automatic2
Manual1
Manual2
Manual3
Manual4
FC1~4
1,1,0,1
1,1,1,1
1,1,0,0
1,0,0,0
1,0,1,0
1,0,0,1
VBAR
Full Scale
Input Terminal
0
0
X
X
X
X
220.00A / 2200.0A IVSL/IVSH V.S. SGND
22.000A / 220.00A IVSL/IVSH V.S. SGND
22.000A
220.00A
2200.0A
22000A
ADP V.S. SGND
ADP V.S. SGND
ADP V.S. SGND
ADP V.S. SGND
Note:
1. Connect VBAR to V- will disable the “µ2” / ”m2” symbol on LCD panel.
2. In DC current modes for clampmeter, ES51932 provides Zero function for offset removing.
3. The typical connection of input terminal for automatic1&2 is as the schematic below:
IVSL
IVSH
Vin*
RL
9R
R
Ver. 1.2
9