ES51932
22,000 Counts Auto DMM
Open circuit voltage for Ω
measurement (except 220Ω)
Ω and Continuity
Mode
-0.58
-0.53
-0.48
V
Internal pull-low to V-
current
AC frequency response at 2.2000V
range
Between GND Pin
and SLACDC
±1%
1.5
µA
HZ
40-400
—
—
—
—
—
—
±5%
400-2000
—
100
—
3dB=Full (ADP)
3dB=10k (ADP)
3dB=1k (ADP)
—
10
1
kHz
kHz
kHz
3dB frequency for LPF mode
f3dB
—
Reference voltage temperature
coefficient
100KΩ resister
Between VRH
0℃<TA<70℃
—
—
ppm/℃
TCRF
TPW
75
Minimum pulse width for Peak
Hold feature
4MHz
crystal
100
µs
oscillator
—
—
—
—
-1.0
-10
-1.0
-10
-2.0
20
1.0
10
1.0
10
2.0
20
%
counts
%
counts
%
22nF – 220nF
Capacitance measurement accuracy2
2.2µF
22µF~220mF
counts
Note:
1.Full Scale
2.Above specification from 22nF to 2.2uF ranges are specified under REL mode. The
reading is calibrated to zero by relative function, when input is floating.
Function Description
1. Operating Modes
1.1 Voltage Measurement
A re-configurable voltage divider automatically provides a suitable range in voltage
measurement mode. The following table summarizes the full-scale ranges in each
configuration.
Configuration Full Scale Range Divider Ratio Resistor Connection
Input Pin
VR1
VR2
VR3
VR4
VR5
220.00mV1
2.2000V
22.000V
220.00V2
2200.0V
1
1/10
1/100
1/1000
1/10000
-
V220m V.S. SGND
VR1 V.S. SGND
VR1 V.S. SGND
VR1 V.S. SGND
VR1 V.S. SGND
R2 / (R1+R2)
R3 / (R1+R3)
R4 / (R1+R4)
R5 / (R1+R5)
Note:
1. 220.00mV range is independent and manual mode. The input pin is different from other voltage
ranges.
2. If VST is set to V- level, the auto range mode will start from 220.00V range. 2.2000V – 22.000V
ranges only exist in manual mode.
Ver. 1.2
6
10/05/27