ES51985
6000 Count AUTO DMM
5.4 Operating Timing
ES51985 incorporates a dual slope ADC with four phases: ZI, AZ, INT and DINT.
The timing of each phase is listed below.
*Voltage /ohm/ADP measurement:
Phase
ZI
Time
100ms
100ms
100ms
600ms
AZ
INT
DINT
Ps. In the voltage and ohm measurement with auto mode, if the range is changed,
the internal clock rate will increase ten times and the new measurement cycle
becomes 1/10 times of the original cycle until the range is stable.
*Current mode
Phase
ZI
Time
100ms
100ms
100ms
300ms
AZ
INT
DINT
*Continuity/diode measurement
Phase
ZI
Time
10ms
10ms
10ms
60ms
AZ
INT
DINT
*.Frequency measurement
one measurement cycle = 1.05 sec
ps. In the frequency measurement with auto mode, if the range is changed, the
internal clock rate will increase ten times and the new measurement cycle
becomes 1/10 times of the original cycle until the range is stable.
18