ES51967
6,600 Counts Auto DMM
—
-1.5
1.5
%F.S
2Capacitance measurement
accuracy
6.6nF-66mF range
—
-3
3
digits
Note:
1.Full Scale
2.Above specification from 6.6nF to 6.6uF ranges are specified under REL mode.
Function Description
1. Operating Modes
1.1 Voltage Measurement
A re-configurable voltage divider automatically provides a suitable range in voltage
measurement mode. The following table summarizes the full-scale ranges in each
configuration.
Configuration
VR1
Full Scale Range
660.0mV1
6.600V
Divider Ratio
1
Resister Connection
VR1 (10MΩ)
VR2 (1.111MΩ)
VR3 (101KΩ)
VR4 (10KΩ)
VR2
VR3
VR4
VR5
1/10
1/100
1/1000
1/10000
66.00V
660.0V2
1000V
VR5 (1KΩ)
Note:
1. 660.0mV range only exists in manual mode when VST pin is kept floating or DGND.
2. If VST is set to V- level, the auto range mode will start from 660.0V range. 660.0mV – 66.00V
ranges only exist in manual mode.
1.1.1 CE selection
ES51967 has a CE selection feature archived by configuring a CESEL pin. If pin
CESEL is connected to V-, ES51967 will have a 610.0V overflow level in DC and AC
660.0V ranges. And the 1000V range will not occur anymore. If pin CESEL is
connected to DGND or floating, the ES51967 will have a 1010V overflow level in DC
1000V range and have a 1010V overflow level in AC 1000V range. The meter will
Show OL when the measuring signal reaches the overflow level. The configuration of
CESEL is listed below.
CESEL
DGND or floating
V-
DCV
1010V
610.0V
ACV
1010V
610.0V
V2.6
6