欢迎访问ic37.com |
会员登录 免费注册
发布采购

ES51967 参数 Datasheet PDF下载

ES51967图片预览
型号: ES51967
PDF下载: 下载PDF文件 查看货源
内容描述: 6600计数自动DMM [6,600 Counts Auto DMM]
分类和应用:
文件页数/大小: 30 页 / 542 K
品牌: CYRUSTEK [ Cyrustek corporation ]
 浏览型号ES51967的Datasheet PDF文件第5页浏览型号ES51967的Datasheet PDF文件第6页浏览型号ES51967的Datasheet PDF文件第7页浏览型号ES51967的Datasheet PDF文件第8页浏览型号ES51967的Datasheet PDF文件第10页浏览型号ES51967的Datasheet PDF文件第11页浏览型号ES51967的Datasheet PDF文件第12页浏览型号ES51967的Datasheet PDF文件第13页  
ES51967  
6,600 Counts Auto DMM  
1.2 Resistance Measurement  
A re-configurable divider automatically provides a suitable full-scale range in resistance  
measurement mode. The following table summarizes the full-scale ranges and the  
reference resistors in each configuration. In 66Mrange, the resolution is 0.1M, if  
SHIFT is floating or connected to DGND. It will be changed to 0.01M, if DSHIFT is  
pulled to V-.  
Configuration  
OR1  
Full Scale Range  
660.0Ω  
Relative Resistor  
OR1  
Equivalent value  
100Ω  
OR2  
6.600KΩ  
VR5  
1KΩ  
OR3  
66.00KΩ  
VR4  
10KΩ  
OR4  
660.0KΩ  
100KΩ  
VR3VR1  
OR5  
OR6  
6.600MΩ  
66.00MΩ  
1MΩ  
10MΩ  
VR2VR1  
VR1  
1.3 Current measurement for multimeter  
ES51967 has 2 automatic and 1 manual current measurement modes for multimeter.  
The following table summarizes the full-scale range of each mode. When ES51967  
operates in the current measurement modes for multimeter, It takes high input from pin  
IVSH or IVSL, low input from pin SGND and reference voltage from pin VR.  
Mode  
FC1~4  
1,1,0,1  
1,1,1,1  
0,0,0,0  
1VASEL 2VBAR  
Full Scale  
Input Terminal  
Automatic1  
Automatic2  
Manual  
1
1
1
1
1
X
660.0µA / 6600µA 3IVSL/IVSH V.S. SGND  
66.00mA / 660.0mA 3IVSL/IVSH V.S. SGND  
66.00A  
IVSH V.S. SGND  
Note:  
1. Connect VASEL to V- will disable “A” and enable “V” units on LCD panel, so in other modes, this pin  
must keep floating.  
2. Connect VBAR to V- will disable the “µ2” / ”m2” symbol on LCD panel, so in the mV range of voltage  
measurement, this pin must keep floating.  
3. The typical connection of input terminal is as the schematic below:  
Vin*  
9R  
IVSL  
IVSH  
100K  
100K  
R
*The voltage results from the current flows through resistors R and 9R. Its input range is up to 3.3Vrms.  
1.4 Current measurement for clampmeter  
ES51967 has 2 automatic and 4 manual current measurement modes for Clampmeter.  
The following table summarizes full-scale range of each mode. When ES51967  
operates in automatic mode1&2, it takes high input from IVSH/IVSL (higher  
range/lower range), low input from SGND and reference voltage from VR. When  
ES51967 operates in the manual mode1~4, It takes high input from ADP, low input  
from SGND and reference voltage from VA+ and VA-.  
V2.6  
9
08/10/06  
 复制成功!