ES51964(6600counts)
Dual Display
1.2 Dual display V.S. measurement mode
The main-display always shows reading in main measurement mode, and sub-display shows
reading in sub-measurement mode or special function. The configuration is shown at the
table below.
Main measurement mode Main-display Sub-display default
DC voltage
AC voltage
DC current
DC voltage
AC voltage
DC current
N/A
frequency
N/A
AC current for multimeter AC current
AC current for clampmeter AC current
frequency
frequency
N/A
N/A
Duty cycle
N/A
Resistance mode
Capacitance mode
Frequency mode
Continuity Check
Diode mode
Resistance
Capacitance
Frequency
Resistance
Voltage
N/A
Temperature mode
ADP mode
Temperature
A/D result
N/A
N/A
Note: N/A means that sub-display shows nothing if no push function is enabled.
1.3 Dual display V.S. push function
With dual display, user can observe current measurement result and special function at the
same time. In other single-display DMM, it can not show the current value when some
special function, for example, Relative, Max/Min, is enabled. In order to see the current
measurement value, the enabled function must be canceled.
Push function
Enable Hold
Max/Min
Relative
Main-display
Stop updating
Current value
Relative value1
Modified value2
Sub-display
Stop updating
Max. or Min. of current value
Reference Value
N/A
Zero
1Relative value = Current A/D result – Reference value
2Modified value = Current A/D result – zero offset value.