ES51964(6600counts)
Dual Display
5.3. Mdigit_3 – Mdigit_0 and Sdigit_3 – Sdigit_0
Mdigit_3 – Mdigit_0 is the measurement result of main measurement mode. This result is
shown on main-display of LCD panel. Sdigit_3 – Sdigit_0 is the measurement result of
sub-measurement mode, for example, frequency in AC voltage/current mode or duty cycle
in frequency mode. This result will be shown on sub-display of LCD panel. Mdigit3 and
Sdigit3 is the most significant digit, and Mdigit0 and Sdigit0 is the least significant digit.
Digit
Code
0
1
2
3
4
5
6
7
8
9
0110000
0110001
0110010
0110011
0110100
0110101
0110110
0110111
0111000
0111001
5.4. STATUS
The format of this package is shown below. The TF field is meaningful only when the
Function packet indicates Temperature mode. In Temperature mode, the TF is 1 if the unit is
°C and is 0 of the unit is °F. BATT field is one when battery low condition is true. The
Judge bit is available in duty cycle mode. If the Judge bit is 1, the final result should be
calculated by (100-X) % which X is the output data from Sdigit_3~Sdigit0. If the Judge bit
is 0, then the final data is X %. The VBAR will be 1 only when the VBAR pin is connected
to V-.
0
1
1
TF
BIT 3
BATT
BIT 2
Judge
BIT1
VBAR
BIT 0
BIT6
BIT 5
BIT 4