ES51964(6600counts)
Dual Display
Push > 2s
Push < 2s
Push < 2s
Normal mode
(Auto/Manual)
HOLD mode
6sec past
(Manual)
Delay State
Push > 2s
Push < 2s
(Auto/Manual)
4.2. Range
RANGE pin switches to and from automatic and manual mode, and while in manual mode,
changes the full-scale range. The following figure shows the state transition.
1 push < 1 sec
Automatic
Mode
Manual
Mode
1 push > 1 sec
Range up
1 push < 1 sec
Measurement Mode
V(DC/AC)
Auto
Manual
Control Range
Initial Range
6.600V
VRi → VRi + 1,
VR5 → VR1
R1 → R2,
R2 → R1
VR2 – VR5
660.0mV – 1000V
Auto µA(DC/AC)
R1 – R2
660.0µA – 6600µA
660.0µA
R1 → R2
R2 → R1
fixed
R1 → R2
R1-R2
fixed
Auto mA(AC/DC)
66 A(DC/AC)
66.00mA – 660.0mA
66.00A
66.00mA
66.00A
66.00A
Auto 66A/660A
(DC/AC)
R1-R2
66.00A – 660.0A
R2 → R1
R1 → R2
R2 → R1
Auto 660A/6600A
(DC/AC)
R1-R2
660.0A – 6600A
6.600nF– 66.00mF
660.0Ω – 66.00MΩ
660.0Ω – 66.00KΩ
660.0A
6.600nF
660.0Ω
660.0Ω
Ci → Ci + 1,
C8 → C1
Capacitance
C1 – C8
ORi → ORi + 1,
OR6 → OR1
Ω
OR1 – OR6
OR1 – OR3
ORi → ORi + 1,
OR3 → OR1
fixed
Ω (Clamp)
Continuity
Diode
fixed
fixed
660.0Ω
6.600V
660.0Ω
6.600V
fixed
FRi → FRi + 1
FR7 → FR1
Frequency
FR1 – FR7
66.00Hz – 66.00MHz
66.00Hz
Note: Pushing RANGE resets all existing special modes.
4.3. KEY
See Section “Measurement Mode Switching” for the function of this pin.
v2.5
17
09/04/14