Measurement parameters and measurement conditions  
Speed and accuracy  
FAST/MED/SLOW/SLOW2  
Measurement speed  
Rs  
Rp  
Ls  
Lp  
Cs  
Cp  
D
Equivalent series resistance (ESR)  
Equivalent parallel resistance  
Equivalent series inductance  
Equivalent parallel inductance  
Equivalent series capacitance  
Equivalent parallel capacitance  
Loss factor tan δ  
Z
Y
θ
X
G
B
Q
Rdc  
Impedance  
Admittance  
Phase angle  
Reactance  
Conductance  
Susceptance  
Q-factor  
Averaging  
Valid setting range: 1 to 256 (in steps of 1)  
Z: 0.05% rdg. θ: 0.03° (representative value)  
Basic accuracy  
Guaranteed  
accuracy range  
Measurement  
parameters  
1 mΩ to 200 MΩ (impedance)  
Guaranteed  
accuracy period  
1 year  
σ
ε
Conductivity  
Permittivity  
DC resistance  
Warm-up time  
60 minutes  
Terminal structure  
4-terminal structure  
Rs ±(0.00 m to 9.99999 G  
Rp ±(0.00 m to 9.99999 G  
)
)
Ω
Ω
Z
Y
θ
X
G
B
Q
Rdc  
0.00 m to 9.99999 G  
Ω
0.000 n to 9.99999 GS  
±(0.000° to 999.999°)  
±(0.00 m to 9.99999 G  
±(0.000 n to 9.99999 GS)  
±(0.000 n to 9.99999 GS)  
±(0.00 to 9999.99)  
Ls ±(0.00000 μ to 9.99999 GH)  
Lp ±(0.00000 μ to 9.99999 GH)  
Cs ±(0.0000 p to 9.99999 GF)  
Cp ±(0.0000 p to 9.99999 GF)  
Supplementary functionality  
Uses a specific signal to time the start of measurement.  
[Trigger types]  
Internal trigger: Automatically generates a trigger signal  
internally to repeat measurement.  
External trigger: Allows you to control the instrument s  
measurement operation by inputting a trigger signal from  
an external device (trigger sources: manual, communica-  
)
Ω
Display range  
D
±(0.00000 to 9.99999)  
Δ% ±(0.000% to 999.999%  
)
,
±(0.00 m to 9.99999 G  
)
Ω
σ
ε
±(0.00000 to 999.999 G)  
±(0.00000 to 999.999 G)  
tions commands, EXT. I/O)  
[Trigger delay]  
Sets the delay time from trigger input to measurement.  
Setting range: 0.0000 s to 9.9999 s  
[Trigger synchronous output]  
.
1 mΩ to 200 MΩ  
Measurable range  
Output impedance  
Range  
Trigger function  
Normal mode: 100 Ω, Low impedance high accuracy mode: 10Ω  
4 Hz to 8 MHz  
Outputs the measurement signal after trigger input and ap-  
plies it to the sample during measurement only.  
Allows you to set a wait time until data is acquired.  
Setting range: 0.0000 s to 9.9999 s  
4.00 Hz to 999.99 Hz …………… 10 mHz steps  
1.0000 kHz to 9.9999 kHz …… 100 mHz steps  
10.000 kHz to 99.999 kHz …… 1 Hz steps  
100.00 kHz to 999.99 kHz …… 10 Hz steps  
1.0000 MHz to 8.0000 MHz …… 100 Hz steps  
Measurement  
Resolution  
frequency  
[Open/short compensation]  
[Load compensation] Number of sets of compensation condi-  
tions: Up to 5  
[Cable length compensation] Cable length settings: 0 m, 1 m,  
2 m, 4 m  
[Correlation compensation] Compensation of display  
values based on user-input compensation coefficient  
Accuracy  
±0.01% of setting or less  
Compensation  
function  
[Normal mode]  
Measurement  
4 Hz to 1.0000 MHz: 10 mV to 5 V (maximum 50 mA)  
1.0001 MHz to 8 MHz: 10 mV to 1 V (maximum 10mA)  
[Low impedance high accuracy mode]  
signal level  
[V mode]  
Range  
[4-terminal contact check]  
Performs a contact (disconnection) check between HCUR and  
HPOT and between LCUR and LPOT.  
[High-Z reject function]  
Detection of OPEN state during 2-terminal measurement.  
[CV mode]  
4 Hz to 1.0000 MHz: 10 mV to 1 V (maximum 100 mA)  
Contact check  
10 mV to 1.000 V …… 1 mV steps  
Resolution  
[Normal mode]  
4 Hz to 1.0000 MHz: 10 μA to 50 mA (maximum 5 V)  
1.0001 MHz to 8 MHz: 10 μA to 1 mA (maximum 1 V)  
[Low impedance high accuracy mode]  
Measurement  
signal level  
[CC mode]  
Range  
Recording and interface  
Measurement result items (maximum 32000 items) can  
4 Hz to 1.0000 MHz: 10 μA to 100 mA (maximum 1 V)  
be saved to the instrument.  
Memory can be read using communications commands  
or a USB flash drive.  
Memory function  
10 μA steps  
Resolution  
Monitor voltage range: 0.000 V to 5.000 V  
Monitor current range: 0.000 mA to 100.0 mA  
Monitor function  
Panel save and load Measurement conditions: Up to 60  
Compensation values: Up to 128  
functions  
DC resistance  
measurement  
EXT. I/O( HANDLER) ,USB, USB flash drive, LAN,  
GP-IB, RS-232C  
Measurement signal level: Fixed at 1 V  
Interfaces  
Generating range: DC voltage 0 V to 2.50 V (10mV resolution)  
In low Z high accuracy mode: 0 V to 1 V (10 mV resolution)  
[Output from EXT. I/O connector]  
Generates BCD output for the No.1 and No.3 parameter  
measured values.  
DC bias measurement  
BCD output  
*Input and output signals are set to BCD mode (selection  
with judgment output).  
Measurement modes  
Display and sound  
LCR mode: Measurement using a single set of conditions.  
Measurement modes Continuous measurement mode: Continuous measurement  
using previously saved conditions  
Lock operation of the instrument using the touch screen.  
Key lock function  
Unlock by entering a passcode.  
Beep tone  
Enable or disable for judgment results and key operation.  
LCD display on/off  
Off: The display turns off 10 sec. after the touch panel is  
last touched.  
LCR mode  
Display settings  
Bin measurement: 10 categories for 2 measurement parameters  
Judgment method: Set as absolute values, percentage, or  
deviation percentage  
Display  
5.7-inch color TFT with touch panel  
Other  
Operating  
temperature and  
humidity  
Measurements  
Comparator measurement: Hi, IN, and Lo judgments for  
2 parameters  
Judgment method: Set as absolute values, percentage, or de-  
viation percentage  
0°C to 40°C (32°F to 104°F), 80% RH, non-condensing  
Storage  
temperature and  
humidity  
Zoom display function: Enlarged display of measured values  
Number of display digits setting: Allows you to set the number  
of display digits for measured values for each measurement  
parameter. (Valid setting range: 3 to 6 digits)  
-10°C to 50°C (14°F to 122°F), 80% RH, non-condensing  
Display  
Operating  
environment  
Indoors, Pollution Degree 2, altitude up to 2000 m (6562-ft.)  
Power supply and  
maximum rated  
power  
100 V AC to 240 V AC (50/60 Hz), 50 VA  
Continuous measurement mode  
Performs continuous measurement using measurement condi-  
tions that have been saved using the panel save function.  
Measurement is started by an external trigger (any of the three  
Dielectric strength  
Standards compliance  
1.62 kV AC for 1 min. between power line and ground line  
Measurements  
EMC: EN 61326, EN 61000  
Safety: EN 61010  
types described below)  
Approx. 330 W × 119 H × 230 D mm (12.99 W × 4.69  
H × 9.06 D in) , approx. 4.2 kg (148.1 oz.)  
Maximum number  
60  
Dimensions and Mass  
of measurements  
Power cord ×1, Instruction manual ×1,  
Accessories  
LCR application disc (Communications user manual) ×1  
 复制成功!