Access an extensive range of interfaces in all model variants  
EXT. I/O  
EXT. I/O allows you to output the measurement complete signal and judgment results signal and to control the instrument by inputting signals such as  
a measurement trigger signal. All signal lines are isolated from the instrument’s measurement and control circuitry for maximum noise resistance.  
Example of EXT I/O timing (LCR mode)  
t7  
t8  
TRIG  
(Start of measurement signal)  
t3  
t4  
t1  
t2  
INDEX  
(Analog end of measurement signal)  
t5  
EOM  
(End of measurement signal)  
PARA -x-x, BINx  
(Comparator and BIN result output)*  
JUDGE RESET  
: ON  
Last judgment  
result  
Judgment result  
Judgment result  
Reset  
ERR  
(Error output)  
JUDGE RESET  
: OFF  
Last judgment  
result  
(Acquires the offset value)  
CALIB  
t6  
LD_VALID (Panel load execution)  
LD0 to LD6  
(Panel number)  
*: PARAx-HI, PARAx-IN, PARAx-LO, AND, BINx, OUT_OF_BINS  
t1: From Comparator, BIN Judgement Result to EOM (LO): Setting value for delay time *1 (Settable range: 0.0000 s to 0.9999 s) ; 40 μs  
t2: From EOM width (LO) to TRIG (LO): Minimum time from end of measurement to next trigger *2 ; 400 μs  
t3: From TRIG (LO) to INDEX (HI): Time from trigger to circuit response *3 ; 400 μs  
t4: INDEX width (HI): Analog measurement time (=Minimum chuck time), switching chuck with INDEX (LO) is possible *4 ; 1 ms  
t5: EOM width (HI): Measurement time *4 ; 1.5 ms  
t6: From TRIG width (LO) to LD-VALID (HI), CALIB (HI): Time to panel load execution and DC adjustment request signal detection: at least t3  
t7: Trigger pulse width (LO time) ; At least 100 μs  
t8: Trigger off (HI time) ; At least 100 μs  
*1. There is an apporoximate error of 100 μs in the delay time entered for Judgement Result EOM for the setting value.  
t1 is the reference value for when the setting value is 0.0000 s.  
*2. t2 is the reference value for when trigger input for during measurement is disabled.  
*3. Additional time is required when loading panel numbers using the panel load function.  
*4. Reference value for Measurement frequency: 1 kHz, Measurement speed: FAST, Range: HOLD