11  
IM3536 connector signal assignment (LCR mode operation)  
EXT. I/O signal list  
Input signals  
TRIG  
: External trigger  
LD0 to LD6  
LD_VALID  
C1  
: Select panel number  
: Execute panel load  
: During BCD output, toggle between  
high-order and low-order digits  
: During BCD output, toggle between  
the No. 1 and No. 3 parameters  
: DC adjustment request  
C2  
CALIB  
Output signals  
EOM  
: End of measurement  
INDEX  
ERR  
ISO_5V  
ISO_COM  
: End of capture  
: Measurement error output  
: Isolated 5V power output  
: Isolated common signal ground  
Output signals (common signal line)  
PARAx-HI, PARAx-IN,  
PARAx-LO (x=1,3),  
AND  
: Comparator judgment  
result output  
Signal assignment is different during continuous measurement mode.  
Signal logic is 0 V to 0.9 V for LO level and 5 V to 24 V for HI level.  
BIN1 to BIN10,  
: BIN judgment result output  
: BCD output signal  
Connectors  
OUT_OF_BINS  
Connectors to use (unit side) : 37-pin D- sub female connector with #4-40 inch screws  
Compliant connectors  
: DC-37P-ULR (solder type) and DCSP-JB37PR (pressure  
weld type)  
For information on where to obtain connectors,  
consult your nearest HIOKI distributor.  
D1-0 to D1-3  
D2-0 to D2-3  
D3-0 to D3-3  
D4-0 to D4-3  
Electrical specifications  
Input signals  
Input type  
Input asserted (on) voltage  
Optocoupler-isolated, non-voltage contact inputs (current sink, active-low)  
0.9 V or less  
Input de-asserted (off) voltage OPEN or 5 V to 24 V  
Input asserted (on) current  
Maximum applied voltage  
Output type  
3 mA/ch  
30 V  
Output signals  
Isolated NPN open-collector outputs (current sink, active-low)  
Maximum load voltage  
Maximum output current  
Residual voltage  
30V  
50 mA/ch  
1 V (10 mA), 1.5 V (50 mA)  
4.5 V to 5.0 V  
100 mA  
Internally isolated Output voltage  
power supply  
Maximum output current  
External power input  
none  
BCD  
LCR mode output signals operate in two modes: judgment mode and BCD mode. In BCD mode, measured values for the  
No. 1 parameter and the No. 3 parameter are output using the BCD signals. *LCR mode only  
C1=HI  
The BCD high-order digit and low-order digit (polarity and ERR informa-  
tion) can be switched with the C1 signal.  
D4-n  
D3-n D2-n  
C1  
D4  
D3  
D2  
D1  
D4-n  
D1-n  
D3-n  
HI (high-order) No. 6 digit data No. 5 digit data No. 4 digit data No. 3 digit data  
C1=LO  
LO (low-order) No. 2 digit data No. 1 digit data  
Polarity  
ERR  
- 8 9 . 4 7 5  
°
D2-n  
Interfaces  
Control the instrument with communication commands from a computer via the USB, LAN, GP-IB, or RS-232C interfaces.  
USB  
LAN  
Connector  
USB Type B receptacle  
Connector  
RJ-45 connector  
Electrical specifications USB2.0 (High Speed)  
Transmission method 10Base-T/100Base-T automatic detection  
Protocol  
TCP/IP  
RS-232C  
GP-IB  
Connector  
D-sub 9-pin connector  
Connector  
24-pin Centronics type connector  
Flow control  
Hardware/Software  
Standard  
IEEE-488.1 1987  
IEEE-488.2 1987  
LF, CR+LF  
Transmission speed  
9600 bps, 19200 bps, 38400 bps, 57600 bps  
Reference standard  
Terminator  
 复制成功!