Agilent 16760A Specifications and Characteristics  
Agilent Technologies 16760A Supplemental Specifications* and Characteristics (continued)  
Synchronous State  
Analysis (continued)  
1.5 Gb/s mode  
1.25 Gb/s mode  
800 Mb/s mode  
400 Mb/s mode  
200 Mb/s mode  
Trigger resources  
3 Patterns on each  
pod evaluated as  
=, , >, <, , ≤  
on one pod; or  
evaluated as =, ≠  
across multiple  
pods; or 1 range  
on each pod  
3 Patterns on each  
pod evaluated as  
=, , >, <, , ≤  
on one pod; or  
evaluated as =, ≠  
across multiple  
pods; or 1 range  
on each pod  
4 Patterns on each  
pod evaluated as  
=, , >, <, , ≤  
on one pod; or  
evaluated as =, ≠  
across multiple  
pods; or 2 ranges  
on each pod  
8 Patterns evaluated 16 Patterns evaluated  
as =, , >, <, , as =, , >, <, , ≤  
4 Ranges evaluated 15 Ranges evaluated  
as in range,  
not in range  
2 Occurrence  
counters  
as in range,  
not in range  
Timers: 2 x (number  
of modules) – 1  
2 Global counters  
1 Occurrence counter  
per sequence level  
4 Flags  
4 Flags  
Arm in  
4 Flags  
Arm in  
4 Flags  
Arm in  
4 Flags  
Arm in  
Arm in  
Trigger actions  
Trigger and  
fill memory  
Trigger and  
fill memory  
Trigger and  
fill memory  
Goto  
Trigger and  
fill memory  
Goto  
Trigger and  
fill memory  
Trigger and goto  
Store/don’t store  
sample  
Turn default storing  
on/off  
Timer start/stop/  
pause/resume  
Global counter  
increment/reset  
Occurrence counter  
reset  
Flag set/clear  
*
All specifications noted by an asterisk are the performance standards against which the product is tested.  
[1] In 1.25 Gb/s and 1.5 Gb/s modes, only the even-numbered channels (0, 2, 4, etc.) are acquired.  
[2] The resolution of the hardware used to assign time tags is 4 ns. Times of intermediate states are calculated.  
[3] The choice of probe can limit system performance. Select a probe rated at the speed of the selected mode (or greater) to maintain system bandwidth.  