Wiring For Magnetic Pickup Coil:  
1. Connect a two conductor shielded cable from the pickup to Terminals 1 and 2. Connect the  
shield to Terminal 4 (see Figure 2). NOTE: It is also preferable that the pickup coil be grounded  
to the shield.  
2. Connect a power source, which must be within 8-30 VDC, to Terminals 7 and 8. The Plus  
voltage is connected to Terminal 8 and the Return to Terminal 7(see Figure 2).  
3. Connect the Pulse output wire to Terminal 3, and connect the Pulse common to Terminal 4 (see  
Figure 2).  
4. Set Jumper HD2 (see Figure 3) to the MAG position.  
5. Configure Jumper HD1 (see Figure 3) to the desired pulse output as follows:  
Pulse Output Type  
Open Collector  
HD1 Jumper Configuration  
JP2  
Square Wave to Vin  
TTL/CMOS  
JP1 and JP2  
JP1, JP2 and JP3  
JP1, JP2 and JP4  
0-10 VDC Square Wave  
5. Install pickup (e.g. in meter) and provide power to PA1001A. Determine if pulse output  
is being generated in the absence of motion. Adjust the MAG Trigger (R19) (see Figure  
3) clock wise until the false pulsation has stopped.  
PA1001A – 7/02  
Page 6 of 11  