AC-DC Power Supplies
ALM85 Series
EMC: Immunity
Phenomenon
Standard
Test Level
Criteria
Notes & Conditions
ESD
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
3
A
A
A
A
A
15 kV air/ 8 kV contact
80-2700 MHz. IEC60601-1-2 Ed.4 at other frequencies
Radiated
EFT/Burst
Surge
Conducted
Magnetic Fields
10 V/m
3
Installation Class 3
10 V
4
A
Dip: 30% 500 ms
Dip: 60% 200 ms
A/B
A/B
High Line/Low Line
High Line/Low Line
Dip: 80% 5000 ms
Int: 100% 10 ms
Int: 100% 20 ms
B
A
B
B
A
A
A
A
B
EN61000-4-11
EN60601-1-2
Int: 100% 5000 ms
Dips and Interruptions
Dip: 30% 25 AC Cycles
Dip: 60% 5 AC Cycles
Int: 100% 0.5 AC Cycles
Int: 100% 1.0 AC Cycles
Int: 100% 250 AC Cycles
230 VAC 100% load, 100 VAC 80%
230 VAC 100% load, 100 VAC 30%
230 VAC 100% load, 100 VAC 50%
Safety Approvals
Safety Agency
Safety Standard
UL60950-1 2nd Edition
EN60950-1
Notes & Conditions
UL
TUV
CB
Information Technology
IEC60950-1
CE
LVD
UL
ANSI/AAMI ES 60601-1
CSA C22.2 No. 60601
EN60601-1
CSA
TUV
CB
Medical
IEC60601-1
Mechanical Details
AC Input
IEC 3ꢀ0-C14
-A Versions
35.43 0.79 (900.0 ꢀ0.0)
1.5 (38.4)
12 & 15 V Versions
5.31 (135.0)
1.58 (40.ꢀ)
KPPX-4P
(Non-Locking)
ꢀ.44
ø0.685 (17.4)
(6ꢀ.0)
AC Input
IEC 3ꢀ0-C14
Standard Version
1.46
1.95 0.039
(50.0 10.0)
0.17 (4.4)
0.37 0.01ꢀ
(9.5 0.3)
(37.0)
Pin
Connection
+Vout
1
2
+Vout
3
RTN
RTN
Not Connected
AC Input
IEC 3ꢀ0-C8
Cꢀ-8 Versions
4
1.46
(37.0)
Shell
0.039 (1.0)
19 & 24 V Versions
4.76 (1ꢀ1.0)
3.04 (77.ꢀ)
1.89 (48.0)
Return (-)
1.46
(37.0)
Output (+)
1.58 (40.ꢀ)
Notes
1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except
output cable length.
2. Weight: 0.78 lbs (355 g) approx.
3. For European mains lead order part EU-MAINS-IEC for C14 versions,
or EU-MAINS-8 for C8 versions.
5. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8
for C8 versions.
6. Output connector: 12 V & 15 V: 4 pin power din with pin 1 & 2 - posiitive and pin
3 & 4 - return, equivalent to KPPX-4P (non-locking). 19 V & 24 V: DC barrel jack;
5.5 outer diameter, 2.5 mm inner diameter with centre positive.
4. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8
for C8 versions.
www.xppower.com
14 Feb 18