SCA61T Series  
data transfer out from MISO starts with the falling edge of SCK immediately after the last bit of  
the SPI command is sampled in on the rising edge of SCK  
maximum SPI clock frequency is 500kHz  
maximum data transfer speed for RDAX is 5300 samples per sec / channel  
The SPI command can be either an individual command or a combination of command and data.  
In the case of combined command and data, the input data follows uninterruptedly the SPI  
command and the output data is shifted out parallel with the input data.  
The SPI interface uses an 8-bit instruction (or command) register. The list of commands is given in  
Table below.  
Command  
Name  
Command  
Format  
Description:  
MEAS  
RWTR  
RDSR  
RLOAD  
STX  
00000000  
00001000  
00001010  
00001011  
00001110  
00010000  
Measure mode (normal operation mode after power on)  
Read and write temperature data register  
Read status register  
Reload NV data to memory output register  
Activate Self test for X-channel  
RDAX  
Read X-channel acceleration through SPI  
Measure mode (MEAS) is standard operation mode after power-up. During normal operation, the  
MEAS command is the exit command from Self test.  
Read temperature data register (RWTR) reads temperature data register during normal  
operation without effecting the operation. Temperature data register is updated every 150 µs. The  
load operation is disabled whenever the CSB signal is low, hence CSB must stay high at least 150  
µs prior the RWTR command in order to guarantee correct data. The data transfer is presented in  
Figure below. The data is transferred MSB first. In normal operation, it does not matter what data is  
written into temperature data register during the RWTR command and hence writing all zeros is  
recommended.  
CSB  
0
1
2
3
4
5
6
7
8
9
10  
11  
12  
13  
14  
15  
SCK  
DATA IN  
COMM AND  
MOSI  
M ISO  
0
0
7
6
6
5
5
4
3
2
2
1
1
DATA OUT  
HIGH IM PEDANCE  
7
4
3
Figure 10.  
Command and 8 bit temperature data transmission over the SPI  
Self test for X-channel (STX) activates the self test function for the X-channel (Channel 1). The  
Internal charge pump is activated and a high voltage is applied to the X-channel acceleration  
sensor element electrode. This causes the electrostatic force that deflects the beam of the sensing  
element and simulates the acceleration to the positive direction. The self-test is de-activated by  
giving the MEAS command.  
Read X-channel acceleration (RDAX) accesses the AD converted X-channel acceleration signal  
stored in acceleration data register X.  
During normal operation, acceleration data registers are reloaded every 150 µs. The load  
operation is disabled whenever the CSB signal is low, hence CSB must stay high at least 150 µs  
prior the RDAX command in order to guarantee correct data. Data output is an 11-bit digital word  
that is fed out MSB first and LSB last.  