 浏览型号71M6533H-IGT/F的Datasheet PDF文件第63页浏览型号71M6533H-IGT/F的Datasheet PDF文件第64页浏览型号71M6533H-IGT/F的Datasheet PDF文件第65页浏览型号71M6533H-IGT/F的Datasheet PDF文件第66页浏览型号71M6533H-IGT/F的Datasheet PDF文件第68页浏览型号71M6533H-IGT/F的Datasheet PDF文件第69页浏览型号71M6533H-IGT/F的Datasheet PDF文件第70页浏览型号71M6533H-IGT/F的Datasheet PDF文件第71页  
FDS_6533_6534_004  
71M6533/71M6534 Data Sheet  
Since TC1 and TC2 are given in µV/°C and µV/°C2, respectively, the value of the VREF voltage  
(1.195V) has to be taken into account when transitioning to PPM/°C and PPM/°C2. This means  
that PPMC = 26.84*TC1/1.195 and PPMC2 = 1374*TC2/1.195).  
3.5.2 Temperature Compensation for VREF  
The bandgap temperature is used to digitally compensate the power outputs for the temperature depen-  
dence of VREF, using the CE register GAIN_ADJ (internal temperature compensation). Since the band  
gap amplifier is chopper-stabilized via the CHOP_E bits, the most significant long-term drift mechanism in  
the voltage reference is removed.  
In internal temperature compensation mode, the CE applies the following formula to determine the  
GAIN_ADJ value. In this formula TEMP_X is the deviation from nominal or calibration temperature ex-  
pressed in multiples of 0.1 °C:  
TEMP _ X PPMC TEMP _ X 2 PPMC2  
GAIN _ ADJ =16385+  
+
214  
223  
3.5.3 System Temperature Compensation  
In a production electricity meter, the 71M6533 and 71M6534 is not the only component contributing to  
temperature dependency. A whole range of components (e.g. current transformers, resistor dividers,  
power sources, filter capacitors) will contribute temperature effects.  
Since the output of the on-chip temperature sensor is accessible to the MPU, temperature compensation  
mechanisms with great flexibility are possible. MPU access to GAIN_ADJ permits a system-wide temper-  
ature correction over the entire meter rather than local to the chip.  
3.5.4 Temperature Compensation for the RTC  
In order to obtain accurate readings from the RTC, the following calibration procedures are recommend-  
ed:  
1. At the time of meter calibration, the crystal oscillator is calibrated using the RTCA_ADJ register in I/O  
RAM to be as close to 32768 Hz as possible. The recommended procedure is to connect a high-  
precision frequency counter to the TMUXOUT pin and select 0x11 for TMUX[4:0]. This will generate a  
4-second pulse at TMUXOUT that can be used to trim RTCA_ADJ to the best value.  
2. When the meter is in service, the MPU takes frequent temperature readings. If the temperature cha-  
racteristics of the crystal are known, the temperature readings can be used to modify the settings for  
the I/O RAM registers PREG[16:0] and QREG[1:0] in order to keep the crystal frequency close to  
32768 Hz.  
3. After periods of operation under battery power, the temperature for the time the meter was not po-  
wered can be estimated by averaging the temperatures before and after battery operation. Based on  
this, the overall correction for the RTC time can be calculated and applied to the RTC after main  
power returns to the meter.  
3.6 Connecting LCDs  
The 71M6533 and 71M6534 has an on-chip LCD controller capable of controlling static or multiplexed  
LCDs. Figure 33 shows the basic connection for an LCD.  
v1.1  
© 2007-2009 TERIDIAN Semiconductor Corporation  
67