欢迎访问ic37.com |
会员登录 免费注册
发布采购

GPMP900 参数 Datasheet PDF下载

GPMP900图片预览
型号: GPMP900
PDF下载: 下载PDF文件 查看货源
内容描述: 大功率电源900瓦AC / DC通用输入电源 [High Power Supplies 900 Watt AC/DC Universal Input Power Supply]
分类和应用:
文件页数/大小: 3 页 / 711 K
品牌: SLPOWER [ SL Power Electronics ]
 浏览型号GPMP900的Datasheet PDF文件第2页浏览型号GPMP900的Datasheet PDF文件第3页  
GPMP900 SERIES
High Power Supplies 900 Watt AC/DC Universal Input Power Supply
SINGLE OUTPUT MEDICAL POWER SUPPLY
SUMMARY
g
• 90 -264 Vac nominal input range
• Overvoltage, over temp. and short circuit protection
• Approved to UL, CSA and EN Standards
• Power Good LED
• Low Leakage
• 12 V Auxiliary output
• Medical and commercial approvals
The GPMP Series are universal input AC/DC power supplies with power
factor correction (PFC) with up to 900 Watts ouput power. The GPMP
product incorporates true inhibit, power fail alarm, power good signal and
fan control. The series is approved to EN60601-1 and EN60950-1 to reduce
design-in time and end system compliance costs
All specifications are typical at nominal input, full load at 25
O
C unless otherwise stated
SPECIFICATIONS
OUPUT SPECIFICATIONS
Output Power
Total Regulation
Rise time
Transient response
Temperature Coef-
ficient
Overvoltage protection
Short Circuit protection
block wall
At turn-on
Main output 25%
0.2 A/ms
Integral forced air
900 Watts max.
See table
3.0 s max.
3.5% max. dev. 350
ms
recovery
to 1%
+/-0.03%/oC
125% +/-10%
Yes
INPUT SPECIFICATIONS
Input Voltage range
Input frequency
Input surge current
Ground Leakage current
Input current
Input current
Input fuse
264 Vac 60 Hz
115 Vac
230 Vac
F1
Universal input
100-240 nom (90 - 264 max)
Vac
47 Hz to 63 Hz
40 A max.
160
mA
12.0 A max
6.0 A max
15.0 A
EMC INFORMATION
Conducted emissions
Radiated emissions
Line freq. harmonics
Voltage fluctuations
ESD Air
ESD Contact
Radiated immunity
Fast Transients (EFT)
Line surge immunity
Conducted immunity
Power freq. mag. field
Voltage dip immunity
EN55011
EN55011
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Level B
Level A
Class A
Complies (clause 5b)
8 kV contact
6 kV air
3 V/m
2 kV 5 kHz
1 kV diff /2 kV cm
3 Vrms
3 A/m
Unit will ride through without loss of
output at 75% of rated load with AC
power fail warning.
ENVIRONMENTAL SPECIFICATIONS
Thermal Performance
operating ambient
(see ratings chart)
non-operating
0 - 50
o
C
Fan cooled
Maximum Altitude
Relative Humidity
Vibration (operating)
Vibration (non-operating)
Shock
Shock
operating / non-operating
non-condensing
0.003 g
2
/Hz
0.026 g
2
/Hz
half-sine 20 g
pk
half-sine 40 g
pk
-20
o
C to +70
o
C
-40 to +85
o
C
900 Watts
10,000 ft. / 40,000
ft. max.
5% - 90% RH
1.5 g
rms
3 axes 10
min/axis
5.0 g
rms
3 axes, 1
hr/axis
10 ms, 3 axes, 6
shocks total
10 ms, 3 axes, 6
shocks total
Page 1 of 3