E52
■ Accessories
It is recommended that the following accessories be used for mounting Temperature Sensors.
Accessory
Temperature
range
Mounting example
Page
Compression 600°C max.
Fitting
Mounting with
Compression
Fitting
30
Terminal box
Compression Fitting
PT screw
Protective tubing
Welding
Note: The Compression Fitting is not
of airtight construction. Do not
use the Compression Fitting for
applications in which the
exposure of the sensing object
will cause problems.
Loose Flange 400°C max.
Mounting with
Loose Flange
Note: 1. Use the Loose Flange in
normal atmospheric
Terminal box
pressure. The Loose
Flange is not of airtight
construction.
Loose Flange
Protective tubing
Mounting screw
2. Use the Loose Flange at
400°C max.
3. Do not apply the Loose
Flange to protective tubing
diameters other than the
applicable ones.
■ Lead Wires
Accessory
Specification
Temperature range Page
For resistance ther-
mometers
Standard
Vinyl-covered with twelve 0.18-dia. conductors (0.3 mm thick) and
4.8 mm in outer dia.
−20°C to 70°C
---
Heat resistive Glass-wool-covered with twenty 0.18-dia. conductors (0.5 mm thick) ex- 0°C to 180°C
ternally shielded with stainless steel, 5.2 mm in outer dia.
Sleeve: 0°C to 100°C
For thermocouples
(compensating con-
ductor)
Standard
Vinyl-covered with seven 0.3-dia. conductors (0.5 mm thick) and 3.2 x −20°C to 70°C
32
5.0 mm in outer dia.
Heat resistive Glass-wool-covered with seven 0.3-dia. conductors (0.5 mm thick) exter- 0°C to 150°C
nally shielded with stainless steel, 2.9 x 4.6 mm in outer dia. Sleeve: 0°C to 100°C
3