R Series Intelligent DAQ –  
Data Acquisition and Control with Onboard Processing  
Environmental  
Safety and Compliance  
NI 78xxR devices are intended for indoor use only.  
Safety  
Operating Environment  
NI 781xR.............................................. 0 to 55 ꢀC (tested in  
NI 78xxR devices are designed to meet the requirements of the following  
standards of safety for electrical equipment for measurement, control,  
and laboratory use:  
accordance with IEC-60068-2-1  
and IEC-60068-2-2)  
• IEC 61010-1, EN 61010-1  
• UL 61010-1, CSA 61010-1  
NI 7830R, NI 7831R  
40 or 80 MHz timebase.................. 0 to 55 ꢀC (tested in  
accordance with IEC-60068-2-1  
and IEC-60068-2-2)  
NI 7833R/7841R/7842R/7851R/7852R/7853R/7854R  
40 MHz timebase ........................... 0 to 55 ꢀC (tested in  
accordance with IEC-60068-2-1  
and IEC-60068-2-2)  
80 MHz timebase ........................... 0 to 55 ꢀC except the  
following: 0 to 45 ꢀC when  
Note: For UL and other safety certifications, refer to the product label or  
visit ni.com/certification, search by model number or product line, and  
click the appropriate link in the Certification column.  
Electromagnetic Compatibility  
This product meets the requirements of the following EMC standards  
for electrical equipment for measurement, control, and laboratory use:  
• EN 61326 (IEC 61326): Class A emissions; Basic immunity  
• EN 55011 (CISPR 11): Group 1, Class A emissions  
• AS/NZS CISPR 11: Group 1, Class A emissions  
• FCC 47 CFR Part 15B: Class A emissions  
installed in an NI PXI-1000/B or  
NI PXI-101x (tested in  
accordance with IEC-60068-2-1  
and IEC-60068-2-2)  
• ICES-001: Class A emissions  
Relative humidity range...................... 10 to 90%, noncondensing  
(tested in accordance with  
CE Compliance  
IEC-60068-2-56)  
This product meets the essential requirements of applicable European  
Directives, as amended for CE marking, as follows:  
Altitude................................................ 2,000 m at 25 ꢀC  
ambient temperature  
• 2006/95/EC; Low-Voltage Directive (safety)  
• 2004/108/EC; Electromagnetic Compatibility Directive (EMC)  
Storage Environment  
Ambient temperature range ............... -20 to 70 ꢀC (tested in  
Note: Refer to the Declaration of Conformity (DoC) for this product for  
any additional regulatory compliance information. To obtain the DoC for  
this product, visit ni.com/certification, search by model number or  
product line, and click the appropriate link in the Certification column.  
accordance with IEC-60068-2-1  
and IEC-60068-2-2)  
Relative humidity range...................... 5 to 95%, noncondensing  
(tested in accordance with  
IEC-60068-2-56)  
Environmental Management  
Note: Clean the device with a soft, nonmetallic brush. Make sure that  
the device is completely dry and free from contaminants before returning  
it to service.  
National Instruments is committed to designing and manufacturing  
products in an environmentally responsible manner. NI recognizes that  
eliminating certain hazardous substances from our products is beneficial  
not only to the environment but also to NI customers.  
Shock and Vibration (for NI PXI-78xxR Only)  
Operational shock ............................... 30 g peak, half-sine, 11 ms  
For additional environmental information, refer to the NI and the  
Environment Web page at ni.com/environment. This page contains the  
environmental regulations and directives with which NI complies, as  
well as other environmental information not included in this document.  
pulse (tested in accordance  
with IEC-60068-2-27; test  
profile developed in accordance  
with MIL-PRF-28800F)  
Waste Electrical and Electronic Equipment (WEEE)  
Random vibration  
EU Customers: At the end of their life cycle, all products must be  
sent to a WEEE recycling center. For more information about WEEE  
recycling centers and National Instruments WEEE initiatives,  
visit ni.com/environment/weee.htm.  
Operating ........................................ 5 to 500 Hz, 0.3 grms  
Nonoperating.................................. 5 to 500 Hz, 2.4 grms  
(tested in accordance with  
IEC-60068-2-64; nonoperating test  
profile exceeds the requirements  
of MIL-PRF-28800F, Class 3)  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  