High-Speed Digitizers – Optimized for Automated Test
Trigger System
Intermodule SMC Synchronization
Using NI-TClk (typical)
Modes ................................................. Edge, hysteresis, window,
digital, immediate, software
Skew.................................................... 500 ps, ≤5 ps after
Sources................................................ CH 0, CH 1, TRIG, PFI <0..1>
PXI_Trig <0..6>, PXI Star
manual adjustment
Power Requirements (typical)
Trigger, and Software
+3.3 VDC (A) +5 VDC (A) +12 VDC (mA) -12 VDC (mA) Total Power (W)
Slope ................................................... Rising or falling
NI 5154 PXI
1.7
2.3
1.7
2.3
1.1
2.5
1.8
2.3
1.8
2.3
1.9
2.4
520
214
520
214
500
200
200
0
200
0
210
0
23.25
21.66
23.25
21.66
21.65
22.65
Hysteresis............................................ Fully programmable
High-frequency reject filter................. 50 kHz software-selectable
Low-frequency reject filter ................. 50 kHz software-selectable
Sensitivity
PCI
NI 5153 PXI
PCI
NI 5152 PXI
PCI
CH 0 and CH 1
NI 5154....................................... 15ꢀ FS
NI 5152, NI 5153........................ 10ꢀ FS
TRIG
Environment
Operating temperature ....................... 0 to +55 °C in all NI PXI
NI 5154....................................... 15ꢀ FS
NI 5153....................................... 10ꢀ FS
NI 5152....................................... 5ꢀ FS
Level accuracy
chassis except the following:
0 to +45 °C when installed in an
NI PXI-1000/B or PXI-101x
chassis (meets IEC 60068-2-1
CH 0, CH 1 ...................................... 5ꢀ FS up to 10 MHz
TRIG ................................................ 1 V up to 10 MHz
Time resolution ................................... 5 ps with time-to-digital
converter enabled
and IEC 60068-2-2)
Storage temperature........................... -40 to +71 °C (meets IEC 60068-
2-1 and IEC 60068-2-2)
Relative humidity ................................ 10 to 90ꢀ, noncondensing
(meets IEC 60068-2-56)
Holdoff................................................. From Rearm Time up to
[(232 – 1) x Sample Clock Period]
Calibration
External Trigger Channel (TRIG)
Impedance
Self-calibration.................................... Gain, offset, triggering, and
timing for all input ranges;
NI 5154, NI 5153 ............................ 2.25 kΩ
NI 5152 ........................................... 1 MΩ || 22 pF
Vertical range...................................... 5 V
Coupling
NI 5154, NI 5153 ............................ DC
NI 5152 ........................................... AC, DC
1 MΩ attenuator (NI 5152 only)
External calibration interval................ 2 years
Certification and Compliances
CE Mark Compliance
For detailed specifications, visit ni.com/manuals.
For certifications, marks, and DoCs, visit ni.com/certification.