欢迎访问ic37.com |
会员登录 免费注册
发布采购

HAL1000UT-A 参数 Datasheet PDF下载

HAL1000UT-A图片预览
型号: HAL1000UT-A
PDF下载: 下载PDF文件 查看货源
内容描述: 可编程霍尔开关 [Programmable Hall Switch]
分类和应用: 开关
文件页数/大小: 28 页 / 805 K
品牌: MICRONAS [ MICRONAS ]
 浏览型号HAL1000UT-A的Datasheet PDF文件第6页浏览型号HAL1000UT-A的Datasheet PDF文件第7页浏览型号HAL1000UT-A的Datasheet PDF文件第8页浏览型号HAL1000UT-A的Datasheet PDF文件第9页浏览型号HAL1000UT-A的Datasheet PDF文件第11页浏览型号HAL1000UT-A的Datasheet PDF文件第12页浏览型号HAL1000UT-A的Datasheet PDF文件第13页浏览型号HAL1000UT-A的Datasheet PDF文件第14页  
HAL1000  
DATA SHEET  
2.3. General Calibration Procedure  
Note: The magnetic south pole on the branded side  
generates negative ADC-READOUT values, the  
north polarity positive values.  
For calibration in the system environment, the applica-  
tion kit from Micronas is recommended. It contains the  
hardware for the generation of the serial telegram for  
programming and the corresponding software for the  
input or calculating of the register values.  
Then, set the system to the calibration point where the  
sensor output must be low, press the key “Readout  
B
” and get the second ADC-READOUT value.  
In this section, the programming of the sensor using  
this tool is explained. Please refer to Section 5. on  
page 23 for information about programming without  
this tool.  
ON  
Now, adjust the hysteresis to the desired value. The  
hysteresis is the difference between the switching  
points and suppresses oscillation of the output signal.  
With 100% hysteresis, the sensor will switch low and  
high exactly at the calibration points. A lower value will  
adjust the switching points closer together. Fig. 2–9  
shows an example with 80% hysteresis.  
For the individual calibration of each sensor in the cus-  
tomer‘s application, a two-point adjustment is recom-  
mended (see Fig. 2–9 for an example). When using  
the application kit, the calibration can be done in three  
steps:  
By pressing the key “calibrate and store”, the software  
will calculate the corresponding parameters for Sensi-  
tivity, VOQ, Low-Level, High-Level and store these val-  
ues in the EEPROM.  
Step 1: Input of the registers which need not be  
adjusted individually  
This calibration must be done individually for each  
sensor.  
The magnetic circuit, the magnetic material with its  
temperature characteristics, and the filter frequency,  
are given for this application.  
The sensor is now calibrated for the customer applica-  
tion. However, the programming can be changed  
again and again if necessary.  
Therefore, the values of the following registers should  
be identical for all sensors in the application.  
– FILTER  
(according to the maximum signal frequency)  
The 500 Hz range is recommended for highest  
accuracy.  
V
OUT  
Hysteresis  
– RANGE  
Sensor  
Sensor  
switched off  
switched on  
(according to the maximum magnetic field at the  
sensor position)  
(here 80 %)  
– TC and TCSQ  
(depends on the material of the magnet and the  
other temperature dependencies of the application)  
position  
Calibration points  
Write the appropriate settings into the HAL1000 regis-  
ters.  
Fig. 2–9: Characteristics of a position switch  
Step 3: Locking the Sensor  
Step 2: Calculation of the Sensor Parameters  
The last step is activating the LOCK function with the  
“lock” key. The sensor is now locked and does not  
respond to any programming or reading commands.  
Fig. 2–9 shows the typical characteristics for a contact-  
less switch. There is a mechanical range where the  
sensor must be switched high and where the sensor  
must be switched low.  
Warning: The LOCKR register cannot be reset!  
Set the system to the calibration point where the sen-  
sor output must be high, and press the key “Readout  
B
”. The result is the corresponding ADC-READ-  
OFF  
OUT value.  
10  
March 4, 2004; 6251-528-1DS  
Micronas