Cold-Junction-Compensated K-Thermocouple-  
to-Digital Converter (0°C to +1024°C)  
mounting technique, and the effects of airflow. Use a  
Applications Information  
large ground plane to improve the temperature mea-  
Serial Interface  
The Typical Application Circuit shows the MAX6675  
interfaced with a microcontroller. In this example, the  
MAX6675 processes the reading from the thermocou-  
ple and transmits the data through a serial interface.  
Force CS low and apply a clock signal at SCK to read  
the results at SO. Forcing CS low immediately stops  
any conversion process. Initiate a new conversion  
process by forcing CS high.  
surement accuracy of the MAX6675.  
The accuracy of a thermocouple system can also be  
improved by following these precautions:  
Use the largest wire possible that does not shunt  
heat away from the measurement area.  
If small wire is required, use it only in the region of  
the measurement and use extension wire for the  
region with no temperature gradient.  
Force CS low to output the first bit on the SO pin. A  
complete serial interface read requires 16 clock cycles.  
Read the 16 output bits on the falling edge of the clock.  
The first bit, D15, is a dummy sign bit and is always  
zero. Bits D14D3 contain the converted temperature in  
the order of MSB to LSB. Bit D2 is normally low and  
goes high when the thermocouple input is open. D1 is  
low to provide a device ID for the MAX6675 and bit D0  
is three-state.  
Avoid mechanical stress and vibration, which could  
strain the wires.  
When using long thermocouple wires, use a twisted-  
pair extension wire.  
Avoid steep temperature gradients.  
Try to use the thermocouple wire well within its tem-  
perature rating.  
Use the proper sheathing material in hostile environ-  
ments to protect the thermocouple wire.  
Figure 1a is the serial interface protocol and Figure 1b  
shows the serial interface timing. Figure 2 is the SO out-  
put.  
Use extension wire only at low temperatures and  
only in regions of small gradients.  
Open Thermocouple  
Bit D2 is normally low and goes high if the thermocou-  
ple input is open. In order to allow the operation of the  
open thermocouple detector, T- must be grounded.  
Make the ground connection as close to the GND pin  
as possible.  
Keep an event log and a continuous record of ther-  
mocouple resistance.  
Reducing Effects of Pick-Up Noise  
The input amplifier (A1) is a low-noise amplifier  
designed to enable high-precision input sensing. Keep  
the thermocouple and connecting wires away from  
electrical noise sources.  
Noise Considerations  
The accuracy of the MAX6675 is susceptible to power-  
supply coupled noise. The effects of power-supply  
noise can be minimized by placing a 0.1µF ceramic  
bypass capacitor close to the supply pin of the device.  
Chip Information  
TRANSISTOR COUNT: 6720  
PROCESS: BiCMOS  
Thermal Considerations  
Self-heating degrades the temperature measurement  
accuracy of the MAX6675 in some applications. The  
magnitude of the temperature errors depends on the  
thermal conductivity of the MAX6675 package, the  
_______________________________________________________________________________________  