14 | Keysight | IntegraVision PA2200 Series Power Analyzers - Data Sheet
Overview
Basic accuracy (50/60 Hz)
0.05% of reading
Best power accuracy (50/60 Hz)
Power Channels (Voltage and Current)
0.1% of reading
PA2201A: 2 Channels,
PA2203A: 4 Channels
Voltage Measurement Bandwidth (-3 dB)
2.5 MHz (-3 dB)
Current Measurement Bandwidth (2 A or 50 A Input) 100 kHz (-3 dB)
Current Measurement Bandwidth (External Input)
Maximum Voltage
2.5 MHz (-3 dB)
1000 Vrms (2000 V peak)
Direct input:
Maximum Current
2 Arms (6 A peak) or 50 Arms (100 A peak)
10 Vrms (30 V peak)
External transducer:
Record Size
Maximum 1.5 M points on each waveform simultaneously
Digitizing Speed
Maximum 5 M samples/second at 16 bits on each waveform simultaneously
12.1- inch capacitive multi-touch/gesture enabled display
Display size and type
Power Channels (Voltage and Current)
Voltage
Connector
Safety Banana Plug
Maximum Voltage
Input Impedance
Ranges
1000 Vrms, 2000 Vpeak
2 M Ω || 12.5 pF
5 V, 10 V, 20 V, 50 V, 100 V, 200 V, 500 V, 1000 V
Pluggable Terminal Block, Phoenix Contact PN: 1967469
2 Arms, 6 A peak or 50 Arms, 100 A peak
2 A input: 60 mΩ + 0.10 µH
Current
Directly Connected
Connector
Maximum Current
Input Impedance
50 A input: 6 mΩ + 0.06 µH
Ranges
2 A input: 10 mA, 20 mA, 50 mA, 100 mA, 200 mA, 500 mA,
1 A, 2 A
50 A input: 200 mA, 500 mA, 1 A, 2 A, 5 A, 10 A, 20 A, 50 A
Isolated BNC
External Transducer
Connector
Maximum Input Voltage
Input Impedance
Ranges
10 Vrms, 30 V peak
100 kΩ || 100 pF
50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V, 10 V
Crest Factor
3 at full-scale (Unless in conflict with maximum input ratings)
Maximum Crest Factor of 30 at 10% of full scale
Current Input Protection
Internal current shunt measurement paths are not fused. Current limit protection (fuse or breaker)
and appropriate wire sizing should be provided by the user. Currents in excess of 140% of the
rated rms currents may cause permanent damage to the current measurement shunts. An internal
protection mechanism is provided for the 2A shunt to avoid damage due to mis-wiring, but it
should not be relied upon in situations where higher than 2 Arms currents are expected.
Isolation
Voltage and current terminals are isolated from earth ground to 1000 V CAT II rating. Voltage is
isolated from current with 1000 V CAT II rating