5 | Keysight | N9923A FieldFox RF Vector Network Analyzer - Data Sheet  
FieldFox RF Vector Network Analyzer (continued)  
Description  
Information  
2-port calibration  
(with Option 122)  
Full 12-term error correction. Corrects for directivity, source match, reflection frequency response, load  
match, and transmission frequency response for an N9923A with Option 122, full two-port S-parameters.  
Unknown thru and QSOLT are both available.  
1-port QuickCal  
(with Option 112)  
Corrects for directivity, frequency response, and source match errors, using internal and external standards.  
QuickCal is most accurate for DUTs with 7/16 and Type-N connectors and measurement uncertainties are  
provided for frequencies ≤ 6 GHz. Reduced accuracy for DUTs with 3.5 mm (m), SMA (m), or other male  
coaxial connectors; performance is unspecified. QuickCal is not recommended for DUTs with 3.5 mm (f), SMA  
(f), or other similar female connectors. QuickCal is not applicable to waveguide.  
2-port QuickCal  
(with Option 112 and 122)  
Full 12-term error correction. Corrects for directivity, source match, reflection frequency response, load  
match, and transmission frequency response. QuickCal is most accurate for DUTs with 7/16 and Type-N con-  
nectors and measurement uncertainties are provided for frequencies ≤ 6 GHz. Reduced accuracy for DUTs  
with 3.5 mm (m), SMA (m), or other male coaxial connectors; performance is unspecified. QuickCal is not rec-  
ommended for DUTs with 3.5 mm (f), SMA (f), or other similar female connectors. QuickCal is not applicable to  
waveguide.  
Guided calibration wizard  
FieldFox’s calibration wizard recommends a calibration type and calibration kit based on selected parameters  
and connector types. Alternatively, users can select their own calibration type and calibration kit.  
Interpolated error correction  
With any type of accuracy enhancement applied, interpolated mode recalculates the error coefficients when  
the test frequencies are changed. The number of points can be increased or decreased and the start/stop  
frequencies can be changed, but the resulting frequency span must be a subset of the original calibration  
frequency span.  
Measurement Throughput  
Cycle time data, 1001 points, CalRdy, typical information.  
Measurement speed  
S11:1.75-3.85 GHz1  
S21:1.78-2.06 GHz1  
0.36 ms/point  
0.36 ms/point  
RF source  
LO  
To receivers  
Port 1  
Port 2  
Figure 1: N9923A with Option 122  
1. 1.4 ms/pt; applicable for N9923A with serial number prefix < than MY5607/SG5607/US5607 and N9923A not upgraded with Option N9910HU-500.