Spectrum Analyzer (continued)
Residual Responses
Input terminated, 0 dB attenuation, preamplifier off, RBW ≤ 1 kHz, VBW auto coupled
20 MHz to 3 GHz
> 3 to 6 GHz
–90 dBm (nominal)
–85 dBm (nominal)
Spurious Responses
Input Mixer level –30 dBm
RFsig = RFtune + 417 MHz
–70 dBc (nominal)
–80 dBc (nominal)
RFsig = RFtune + 1.716 GHz
Input Mixer level –10 dBm; First IF Image Response
Rfsig = Rftune – 2 x 0.8346 GHz
(for Rftune 5.7 to 6 GHz)
–50 dBc (nominal)
–80 dBc (nominal)
Sidebands
–60 dBc (nominal) when battery charging, 260 kHz offset
Figure 10
21