Spectrum Analyzer (continued)
Absolute Amplitude Accuracy at 50 MHz
Peak detector, 10 dB attenuation, preamplifier off, RBW < 2 MHz, input signal –5 dBm to –50 dBm, all settings auto–coupled
Specification
Typical
10 minute warm up
90 minute warm up
± 0.4 dB
20 to 30 ºC
± 0.8 dB
± 1.1 dB
± 0.8 dB
-
–10 to 55 ºC
± 0.8 dB
Frequency Response
Relative to 50 MHz, Peak detector, 10 dB attenuation, preamplifier off, RBW = 30 kHz, input signal 0 dBm to –50 dBm, all
settings auto–coupled
Preamp off
20 to 30 ºC
2 to 10 MHz
± 1.1 dB
± 0.9 dB
± 1.3 dB
± 1.5 dB
± 1.0 dB
± 0.6 dB
± 1.1 dB
± 1.5 dB
± 0.5 dB
± 0.3 dB
± 0.5 dB
± 0.5 dB
> 10 MHz to 3 GHz
> 3 to 5 GHz
> 5 to 6 GHz
–10 to 55 ºC
2 to 10 MHz
± 2.0 dB
± 1.5 dB
± 2.0 dB
± 2.6 dB
-
-
-
-
± 1.0 dB
± 0.6 dB
± 1.1 dB
± 1.5 dB
> 10 MHz to 3 GHz
> 3 to 5 GHz
> 5 to 6 GHz
Preamp on
20 to 30 ºC
2 to 10 MHz
-
-
-
-
-
-
-
-
± 0.7 dB
± 0.5 dB
± 0.7 dB
± 0.7 dB
> 10 MHz to 3 GHz
> 3 to 5 GHz
> 5 to 6 GHz
–10 to 55 ºC
2 to 10 MHz
-
-
-
-
-
-
-
-
± 1.2 dB
± 0.8 dB
± 1.3 dB
± 1.7 dB
> 10 MHz to 3 GHz
> 3 to 5 GHz
> 5 to 6 GHz