欢迎访问ic37.com |
会员登录 免费注册
发布采购

N9912A-308 参数 Datasheet PDF下载

N9912A-308图片预览
型号: N9912A-308
PDF下载: 下载PDF文件 查看货源
内容描述: [Handheld Analyzer]
分类和应用:
文件页数/大小: 34 页 / 2701 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号N9912A-308的Datasheet PDF文件第1页浏览型号N9912A-308的Datasheet PDF文件第2页浏览型号N9912A-308的Datasheet PDF文件第3页浏览型号N9912A-308的Datasheet PDF文件第5页浏览型号N9912A-308的Datasheet PDF文件第6页浏览型号N9912A-308的Datasheet PDF文件第7页浏览型号N9912A-308的Datasheet PDF文件第8页浏览型号N9912A-308的Datasheet PDF文件第9页  
Definitions  
All specifications and characteristics apply over a 25 ±5 °C range (unless otherwise stated) and 90 minutes  
after the instrument has been turned on.  
Specification (spec.)  
Warranted performance. Specifications include guardbands to account for the expected statistical  
performance distribution, measurement uncertainties, and changes in performance due to environmental  
conditions. The following conditions must be met:  
-
-
-
FieldFox has been turned on at least 90 minutes  
FieldFox is within its calibration cycle  
Storage or operation at 25 ±5 °C range (unless otherwise stated)  
Typical (typ.)  
Expected performance of an average unit over a 20 °C to 30 °C temperature range after being at ambient  
temperature for two hours, unless otherwise indicated; does not include guardbands. It is not covered by the  
product warranty. The FieldFox must be within its calibration cycle.  
Nominal (nom.)  
A general, descriptive term or design parameter. It is not tested, and not covered by the product warranty.  
Calibration  
The process of measuring known standards to characterize an instruments systematic (repeatable) errors.  
Corrected (residual)  
Indicates performance after error correction (calibration). It is determined by the quality of calibration  
standards and how well "known" they are, plus system repeatability, stability, and noise.  
Uncorrected (raw)  
Indicates instrument performance without error correction. The uncorrected performance affects the stability of  
a calibration.  
4