34 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet
Infiniium S-Series Performance Characteristics (Continued)
Horizontal
Horizontal system: Oscilloscope channels
Main timebase range
Resolution
5 ps/div to 50 s/div
1 ps
Modes
Main, Delayed, Roll (200 ms/div to 200 s/div)
⦁
Reference position
Horizontal position range
Delayed sweep range
Time scale accuracy 1, 8
Oscilloscope channel de-skew range
Intrinsic jitter 6
Continuously adjustable across horizontal display range
0 to ± 500 sec
1 ps/div to current main time scale setting
± (12 ppb initial + 75 ppb/year aging)
–1 ms to +1 ms
Acquired time range or
delta-time interval
< 1 us (100 ns/div)
10 us (1 us/div)
100 us (10 us/div)
1 ms (100 us/div)
10 ms ( 1 ms/div)
100 fs rms
Internal reference
External reference
(Sample clock jitter)
100 fs rms
123 fs rms
138 fs rms
145 fs rms
200 fs rms
100 fs rms
123 fs rms
138 fs rms
145 fs rms
145 fs rms
Inter-channel intrinsic jitter 3
Inter-channel skew drift 3, 7
Jitter measurement floor 2
– Time interval error
< 500 fs rms
2
NoiseFloor
SlewRate
2
+ Intrinsic Jitter
(
)
– Period jitter
2
NoiseFloor
SlewRate
2
2
3
+ Intrinsic Jitter
(
)
– Cycle-cycle/N-cycle jitter
Inter-channel jitter measurement floor 2, 3, 5
2
NoiseFloor
SlewRate
2
+ Intrinsic Jitter
(
)
2
2
2
Time Interval
Time Interval
Interchannel
+
+
Error Edge1
Error Edge2
(
)
(
)
Intrinsic Jitter
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 5 °C from firmware calibration
temperature.
2. Sample rate at maximum. Noise and slew rate determined at fixed-voltage measurement threshold, near middle of signal. Displayed signal not vertically
clipped. Slew rate of sine wave = (peak signal amplitude)
2 π f, slew rate of fast step ~= (10 to 90% rise time).
⦁
⦁ ⦁
3. Intra-channel = both edges on the same channel, Inter-channel = two edges on different channels. Time Interval Error(Edge1) = time-interval error
measurement floor of first edge, Time Interval Error(Edge2) = time-interval error measurement floor of second edge.
4. Reading is the displayed Delta Time Measurement Accuracy measurement value. Do not double the listed Time Scale Accuracy value in Delta Time
Measurement Accuracy formula.
5. Scope channels and signal interconnect de-skewed prior to measurement.
6. External timebase reference values measured using a Wenzel 501-04608A 10 MHz reference. Intrinsic jitter value depends on acquisition time range for
Time Interval Error formula and depends on delta-time between edges for all two-edge formulas.
7. Skew between channels caused by ± 5 degrees C temperature change.
8. Initial = immediately after factory or user calibration.