32 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet
Infiniium S-Series Performance Characteristics
Vertical
DSO/MSO models
Vertical - scope channels
Input channels
S-054A
S-104A
S-204A
S-254A
S-404A
S-604A
S-804A
DSO models - 4 analog
MSO models - 4 analog + 16 digital
50 Ω 1 500 MHz
1 MΩ 500 MHz
Analog bandwidth (–3 dB)
1 GHz
500 MHz
2 GHz
500 MHz
2.5 GHz
500 MHz
4 GHz
500 MHz
6 GHz 6
500 MHz
8 GHz 6
500 MHz
Vertical resolution 2, 3
Typical rise time/fall time
10 to 90% 4
Typical rise time/fall time
20 to 80% 5
10 bits, up to 12 bits with high-resolution mode
860 ps
620 ps
8.1
430 ps
310 ps
7.8
215 ps
155 ps
7.5
172 ps
124 ps
7.4
107.5 ps
77.5 ps
7.2
71.7 ps
51.7 ps
6.8
53.8 ps
33.8 ps
6.4
ENOB (typical)
Input impedance 1
50 Ω: ± 3.5% (typical ± 1% at 25 °C)
1 MΩ: ± 1% (14 pF typical)
50 Ω: 1 mV/div to 1 V/div
1 MΩ: 1 mV/div to 5 V/div
50 Ω: DC
Input sensitivity 3
Input coupling
1 MΩ: AC (> 11 Hz), DC
Bandwidth limit filters (analog)
DSP bandwidth limit filters
Channel-to-channel isolation
20 MHz, 200 MHz for both 50 Ω and 1 MΩ paths
Increments of 500 MHz up to rated scope bandwidth
DC to 100 MHz: 50 dB
100 MHz to 1 GHz: 40 dB
> 1 GHz: 30 dB
DC gain accuracy 1, 2, 3
Max input voltage 1
Offset range
± 1% of full scale at full resolution (typical 5 mV to 1 V per division)
± 2% of full scale at full resolution 1
50 Ω: ± 5 V
1 MΩ: 300 Vrms or DC and ± 400 Vpp (DC+AC)
1 MΩ
50 Ω
Vertical sensitivity
All vertical ranges
Available offset
± 12 divisions or ± 4 V,
whichever is smallest
Vertical sensitivity
< 10 mV/div
≥ 10 mV/div
≥ 20 mV/div
≥ 100 mV/div
≥ 1 V/div
Available offset
± 2 V
± 5 V
± 10 V
± 20 V
± 100 V
Offset accuracy 1, 3
Dynamic range 7
± 0.1 div ± 2 mV ± 1% of offset setting for offsets < 2 V
± 0.1 div ± 2 mV ± 1.5% of offset setting for offsets ≥ 2 V
50 Ω: ± 4 divisions from center screen
1 mV to 100 mV; 2nd harmonic distortion of –30 dBc
101 mV to 5 V; 2nd harmonic distortion of –20 dBc
1 MΩ: ± 4 divisions from center screen
DC voltage measurement
accuracy 2
Dual cursor: ± [(DC gain accuracy) + (resolution)]
Single cursor: ± [(DC gain accuracy) + (offset accuracy) + (resolution/2)]
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 5 °C from firmware calibration
temperature. Input impedance is valid when V/div scaling is adjusted to show all waveform vertical values within the oscilloscope display.
2. Vertical resolution = 0.4% of full scale(≤ 5 GSa/s) or 0.1% of full scale(10 GSa/s or 20 GSa/s).
3. 50 Ω input: Full scale is defined as 8 vertical divisions. Magnification is used below 2 mV/div, full-scale is defined as 16 mV.
The major scale settings are 5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V. Testing is at maximum sample rate.
1 MΩ input: Full scale is defined as 8 vertical divisions. Magnification is used below 2 mV/div, full-scale is defined as 16 mV.
The major scale settings are 5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V. Testing is at maximum sample rate.
4. Calculation based on Tr = 0.43/BW.
5. Calculation based on Tr = 0.31/BW.
6. 6 GHz and 8 GHz bandwidth supported in 2-channel mode. If all four channels are on, a maximum bandwidth of 4 GHz is supported.
7. For a 10:1 probe on the 1 MΩ input, vertical scaling is multiplied by 10.