41 | Keysight | Infiniium V-Series Oscilloscopes - Data Sheet
Infiniium V-Series Performance Characteristics (Continued)
Measurements and math
Oscilloscope measurements
Number of measurements
Measurement update rate
20 simultaneous measurements (can be made on either main, zoom or gated region)
> 50,000 measurement/sec (one measurement turned on)
> 250,000 measurement/sec (10 measurements turned on)
Measurement access
Drop down measurement list
Multipurpose
Drag-and-drop measurement toolbar
Measurement menu access to all measurements
Front panel button activates up to 10 pre-selected or up to 10 user-defined measurements
Measurement toolbar with common measurement icons that can be dragged and dropped onto the
displayed waveforms
Measurement modes
Statistics
Displays the current, mean, minimum, maximum, range (max-min), standard deviation, count, edge,
direction, meas window, icon x, and icon y
Histograms
Source
Waveform or measurement
Orientation
Vertical (for timing and jitter measurements) or horizontal (noise and amplitude change) modes, regions are
defined using waveform markers
Measurements
Mean, standard deviation, mean ± 1, 2, and 3 sigma, median, mode, peak-to-peak, min, max, total hits,
peak (area of most hits), X scale hits, and X offset hits
Marker modes
Manual markers, track waveform data, track measurements, and delta marker values can be displayed
Waveform measurements
Voltage (analog)
Peak-to-peak, minimum, maximum, average, RMS, amplitude, base, top, overshoot, preshoot, V overshoot,
V preshoot, upper, middle, lower, crossing point voltage, pulse top, pulse base, pulse amplitude
Rise time, fall time, period, frequency, positive width, negative width, duty cycle, Tmin, Tmax, Tvolt,
channel-to-channel delta time, channel-to-channel phase, count pulses, burst width, burst period, burst
interval, setup time, hold time, edge-edge, edge time, slew rate
Time (analog)
Time (digital)
Clock (analog)
Period, frequency, positive width, negative width, duty cycle, delta time
Period, frequency, duty cycle, clock, phase, TIE 1, N-period 1, period-period 1, + width to + width 1, - width
to - width 1, duty cycle- to-duty cycle
Setup time, hold time , time interval error 1, unit interval, noise, N-UI, UI-UI, data rate, clock recovery rate,
DDPWS, de-emphasis
Data (analog)
Mixed (analog)
Area, slew rate
Frequency domain (analog)
FFT frequency, FFT magnitude, FFT delta frequency, FFT delta magnitude, peak detect mode, amplitude
modulation
Eye (analog)
Level qualification
Eye height, eye width, eye jitter, crossing %, Q factor, duty cycle distortion
Any channels that are not involved in a measurement can be used to level-qualify all timing measurements
1� Requires EZJIT Plus (Option N5400A) or EZJIT Complete (Option N8823A) software�