Infiniium V-Series Performance Characteristics (Continued)
Acquisition
Acquisition - oscilloscope channels
Maximum real-time sample rate
2 channels
4 channels
Memory depth per channel
Standard
Option 100 (standard on DSA model) 100 Mpts
Option 200
Option 500
Option 01G
V084A
V134A
V164A
V204A
V254A
V334A
80 GSa/s
40 GSa/s
4 channels
50 Mpts
80 GSa/s
40 GSa/s
80 GSa/s
40 GSa/s
80 GSa/s
40 GSa/s
2 channels
100 Mpts
200 Mpts
500 Mpts
1 Gpts
80 GSa/s
40 GSa/s
80 GSa/s
40 GSa/s
200 Mpts
500 Mpts
1 Gpts
1 Gpts
Option 02G
2 Gpts
2 Gpts
Maximum acquired time at highest
real-time resolution
Standard
Option 100 (standard on DSA model) 2.5 ms
Option 200
40 GSa/s
80 GSa/s
1.25 ms
1.25 ms
2.5 ms
5 ms
5 ms
Option 500
Option 01G
Option 02G
12.5 ms
25 ms
50 ms
12.5 ms
12.5 ms
25 ms
Maximum waveform update rate
Sampling modes - oscilloscope channels
Real-time
> 400,000 waveforms per second (when in segmented memory mode)
Successive single shot acquisitions
Real-time with averaging
Real-time with peak detect
Real-time with hi-resolution
Equivalent time
Averages are selectable from 2 to 65,534
80 GSa/s in 2-channel mode, 40 GSa/s in 4-channel mode
Real-time boxcar averaging reduces random noise and increases resolution
Resolution 338 fs
Gaussian magnitude, linear phase
Roll mode
Slow filter roll off while maintaining linear phase
Scrolls sequential waveform points across the display in a right-to-left rolling motion. Works at sample rates up
to 10 MSa/s with a maximum record length of 40 Mpts.
Segmented memory
Captures bursting signals at maximum sample rate without consuming memory during periods of inactivity.
– Maximum time between triggers is 562,950 seconds
– Re-arm time: 2.5 μs
Memory depth:
20 M
50 M
100 M
200 M
500 M
1 G
2 G
Max # of segments:
4096
8192
16384
32768
65536
131072
131072
Filters - Sin(x)/x interpolation
On/off selectable FIR digital filter. Digital signal process adds points between acquired data points to enhance
measurement accuracy and waveform display.
Acquisition - digital channels
Maximum real-time sample rate
Maximum memory depth per channel
Minimum width glitch detection
10 GSa/s with 16 channels, 20 GSa/s with 8 channels
Up to 1 Gpts
50 ps