09 | Keysight | N8834A MultiScope Application - Data Sheet
Performance Characteristics
Control PC requirement Windows 7 64 bit, minimum 4 G RAM (recommended 8 G), 1 G free hard drive
space Infiniium software 5.50 and above
Note: Control PC is not required for Infiniium Series scopes if leader scope is used
as the controller. A control PC must be used if InfiniiVision scopes exist in the
multiscope configuration
Leader and followers
oscilloscope
requirement1
Infiniium software 5.50 and above 2
Channels supported
Analog channel 1 to 4
Note: Digital channels on all MSO models are not supported
Trigger modes 3
Edge
Triggers on a specified slope (rising, falling, or alternating between rising and
falling) and voltage level on any channel or auxiliary trigger
Glitch
Triggers on glitches narrower than the other pulses in your waveform by
specifying a width less than your narrowest pulse and a polarity
Note: All other trigger modes are not supported
1. Front panel control is disabled during MultiScope operation. Oscilloscope control is handled by the Infiniium
Offline software on PC or the leader scope.
2. Software 4.30 or above requires Windows 7. N2753A Infiniium Windows XP to 7 OS upgrade kit (oscilloscope
already has M890 motherboard). N2754A Infiniium Windows XP to 7 OS and M890 motherboard upgrade kit
(oscilloscope without M890 motherboard). Verify the M890 motherboard using the procedure found in the
Windows 7 upgrade kit data sheet with the publication number 5990-8569EN.
3. If leader scope is used as the controller, all trigger modes supported on the leader scope will be available in
the multiscope configuration.