05 | Keysight | N8834A MultiScope Application - Data Sheet  
Calibration Methods  
There are three calibration methods available with the N8834A  
MultiScope software. Each calibration method provides a  
different level of channel-to-channel skew value. Manual  
calibration is available for users who are interested in viewing  
multi-channel signals but not concerned about the skew factor  
across channels. External trigger helps users achieve 100 ps  
channel-to-channel skew. The most accurate measurements can  
be achieved with the drift correction method with skew in the  
range of 200 fs.  
3. Precision  
The horizontal position of channels within different oscilloscope  
frames can drift with temperature (inter-oscilloscope drift). The  
drift correction synchronization process corrects for this drift by  
monitoring the inter-oscilloscope drift on a separate, dedicated  
synchronization channel. The idea is that any “time corrections”  
that keep the synchronization channels aligned will also keep  
the signals aligned. Precision calibration will consume one of the  
channels on each oscilloscope and is only available on 90000 X-,  
90000 Q-, V-Series, and Z-Series oscilloscopes. On the Infiniium  
90000 X- and V-Series, a maximum of two oscilloscopes can  
be connected to perform precision calibration. A sync port is  
available for precision calibration on 90000 Q- and Z-Series.  
1. Manual  
Manual calibration requires no cabling and users can achieve  
skew correction by manually dialing in the right skew values  
between the channels. InfiniiVision scopes supports only manual  
calibration.  
2. Basic  
The trigger out process provides the most basic level of inter-  
oscilloscope time synchronization. This process connects the  
trigger output signal (Trig Out) from the leader oscilloscope  
through each follower oscilloscope (Aux Trig) in a daisy chain  
configuration. The followers are each armed in descending order  
from furthest from the leader (F{N}) to nearest to the leader (F{1}).  
The leader’s trigger is armed last. Once the leader triggers, each  
follower will trigger in turn, down through the chain.