39 | Keysight | Infiniium Z-Series Oscilloscopes - Data Sheet  
Specifications (Continued)  
Hardware trigger (Continued)  
Timeout  
Trigger when a channel stays high, low, or unchanged for too long. Timeout setting: from 250 ps to 10 s  
Pattern/pulse range  
Triggers when a specified logical combination of the channels is entered, exited, present for a specified period of  
time or is within a specified time range or times out. Each channel can have a value of High (H), Low (L) or Don’t care  
(X)  
State  
Window  
Pattern trigger clocked by the rising, falling or alternating between rising and falling edge of one channel  
Triggers on an event associated with a window defined by two-user adjustable thresholds. Event can be window  
“entered,” “exited,” “inside (time qualified),” or “outside (time qualified)” voltage range. Trigger point can be “cross  
window boundary” or “time out.” Time qualify range: from 250 ps to 10 s  
Video  
Triggers from negative sync composite video, field 1, field 2, or alternating fields for interlaced systems, any field,  
specific line, or any line for interlaced or non-interlaced systems. Supports NTSC, PAL-M (525/60), PAL, SECAM  
(625/50), EDTV (480p/60), EDTV (576p/50), HDTV (720p/60), HDTV (720p/50), HDTV (1080i/60), HDTV (1080i/50),  
HDTV (1080p/60), HDTV (1080p/50), HDTV (1080p/30), HDTV (1080p/25), HDTV (1080p/24), and user-defined  
formats  
Trigger sequences  
Three-stage trigger sequences including two-stage hardware (Find event (A) and Trigger event (B)) and one-stage  
InfiniiScan software trigger. Supports all hardware trigger modes except “edge then edge” and “video,” and all  
InfiniiScan software trigger modes. Supports “delay (by time)” and “reset (by time or event)” between two hardware  
sequences. The minimum latency between “find event (A)” and “trigger event (B)” is 3 ns  
Single or multiple channels may be logically qualified with any other trigger mode  
100 ns to 10 s  
Trigger qualification and qualifier  
Trigger holdoff range  
Trigger actions  
Specify an action to occur and the frequency of the action when a trigger condition occurs. Actions include email on  
trigger and execute “multipurpose” user setting  
Software trigger (Requires N5414B InfiniiScan event identification software – Option 009)  
Trigger modes  
Zone qualify  
Software triggers on the user-defined zones on screen. Zones can be specified as either “must intersect” or “must  
not intersect.” Up to eight zones can be defined across multiple channels  
Generic serial  
Software triggers on NRZ-encoded data up to 8.0 Gbps, up to 80-bit pattern. Support multiple clock data recovery  
methods including constant frequency, 1st-order PLL, 2nd-order PLL, explicit clock, explicit 1st-order PLL, explicit  
2nd-order PLL, Fibre Channel, FlexRay receiver, FlexRay transmitter (requires E2688A except for the constant  
frequency clock data recovery mode)  
Measurement limit  
Software triggers on the results of the measurement values. For example, when the “pulse width” measurement is  
turned on, InfiniiScan measurement software trigger triggers on a glitch as narrow as 40 ps. When the “time interval  
error (TIE)” is measured, InfiniiScan can trigger on a specific TIE value  
Non-monotonic edge  
Runt  
Software triggers on the non-monotonic edge. The non-monotonic edge is specified by setting a hysteresis value  
Software triggers on a pulse that crosses one threshold but fails to cross a second threshold before crossing the  
first again. Unlike hardware runt trigger, InfiniiScan runt trigger can be further qualified via a hysteresis value  