09 | Keysight | M9381A PXIe Vector Signal Generator – Data Sheet  
Technical Specifications and Characteristics  
Amplitude (continued)  
Amplitude switching speed  
List mode switching speed 8  
Baseband power level change 9  
ALC off  
Standard, nominal  
Option UNZ, nominal  
≤ 5 ms  
≤ 5 ms  
≤ 5 ms  
≤ 10 µs  
≤ 105 µs  
≤ 105 µs  
ALC on  
Non-list mode switching speed 10  
Baseband power level change 9  
Arbitrary power level change  
≤ 5 ms  
≤ 5 ms  
≤ 250 µs  
≤ 1.5 ms  
List mode  
See frequency specification section for more detail  
Absolute level accuracy in CW mode [ALC on] 11  
Frequency  
< Max power to −20 dBm  
< −20 to −110 dBm  
< −110 to −120 dBm  
< −120 to −130 dBm  
1 MHz to 3 GHz  
± 0.4 dB  
± 0.5 dB  
± 0.7 dB  
± 0.8 dB, nominal  
± 0.15 dB, typical  
± 0.15 dB, typical  
± 0.25 dB, typical  
> 3 to 6 GHz  
± 0.5 dB  
± 0.6 dB  
± 1.0 dB  
± 0.8 dB, nominal  
± 0.15 dB, typical  
± 0.25 dB, typical  
± 0.5 dB, typical  
-120 dBm Output Power Normalized Response  
1
0.75  
0.5  
0.25  
0
-0.25  
-0.5  
-0.75  
-1  
0
1
2
3
4
5
6
Frequency (GHz)  
Figure 7. Output power normalized response at –120 dBm.  
8. Time from trigger input to amplitude settled within 0.2 dB (within the controlled temperature range) or  
within 0.5 dB (at the full temperature range). Carrier frequency is not changing. Measurements made with  
the M9036A embedded controller in an M9018A chassis.  
9. Baseband offset amplitude settled within 0.2 dB. Baseband offset can be adjusted from 0 to –20 dB.  
10. Mean time from IVI command to amplitude settled within 0.2 dB. Carrier frequency is not changing.  
11. Specifications apply within the controlled temperature range. For temperatures outside this range,  