03 | Keysight | M9485A PXIe Multiport Vector Network Analyzer 1 MHz/50 MHz to 9 GHz - Data Sheet
Configuration
This document provides technical specifications for the M9485A PXIe multiport vector
network analyzer. The configuration is shown below
M9300A
M9309A
M9389A
M9340A
M9376A
M9377A
M9378A
M9378B
PXIe frequency reference
PXIe vector network analyzer synthesizer
PXIe vector network analyzer source (includes M9309A, M9310A)
PXIe vector network analyzer RF distributor
PXIe vector network analyzer receiver
PXIe vector network analyzer direct access receiver
PXIe vector network analyzer directional coupler
PXIe vector network analyzer directional coupler with bias-T
Calibration Kits and ECal Modules
This M9485A data sheet also provides technical specifications for the following calibra-
tion kits and ECal modules. For models not listed in this data sheet, please download
the free Uncertainty Calculator from http://www.keysight.com/find/na_calculator to
generate the curves for your calibration kit and enable VNA setup.
85033E
85093C
Calibration kit
Electronic calibration (ECal) module
Definitions
Specification (spec.):
Warranted performance. All specifications apply at 25 ºC (± 5 ºC), unless otherwise
stated, and 45 minutes after the instrument has been turned on. Specifications include
guard bands to account for the expected statistical performance distribution, measure-
ment uncertainties, and changes in performance due to environmental conditions.
Typical (typ.):
Expected performance of an average unit which does not include guardbands.
General characteristics:
A general, descriptive term that does not imply a level of performance.
Boundary Conditions
If the same boundary conditions fall under more than one category in a table, apply the
best value. For example, system dynamic range in page 4 is 139 dB with the following
boundary conditions.
Frequency: 20 MHz
IF bandwidth: 10 Hz