欢迎访问ic37.com |
会员登录 免费注册
发布采购

M9195A-M12 参数 Datasheet PDF下载

M9195A-M12图片预览
型号: M9195A-M12
PDF下载: 下载PDF文件 查看货源
内容描述: [M9195A PXIe Digital Stimulus/Response with PMU: 250 MHz, 16-channel]
分类和应用:
文件页数/大小: 19 页 / 3790 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号M9195A-M12的Datasheet PDF文件第11页浏览型号M9195A-M12的Datasheet PDF文件第12页浏览型号M9195A-M12的Datasheet PDF文件第13页浏览型号M9195A-M12的Datasheet PDF文件第14页浏览型号M9195A-M12的Datasheet PDF文件第15页浏览型号M9195A-M12的Datasheet PDF文件第16页浏览型号M9195A-M12的Datasheet PDF文件第18页浏览型号M9195A-M12的Datasheet PDF文件第19页  
17 | Keysight | M9195A PXIe Digital Stimulus/Response with PMU: 250 MHz, 16-channel - Data Sheet  
Technical specifications and characteristics  
Environmental Characteristics  
Operating and storage conditions  
Temperature  
Operating  
Storage  
0 to 45 °C  
-40 to +70 °C  
Humidity  
Maximum 80% at 40°C (non-condensing)  
Altitude  
3,000 meters, de-rate max temperature by 5 °C above 2,000 meters  
Calibration interval  
Mechanical  
1 year: return to Keysight service center or use Y1252A with N7800A/N7867A  
Operating vibration  
Survival sine vibration  
Survival random vibration  
Transportation shock  
End use handling shock  
Warm-up time  
5-500 Hz: 0.21 g RMS, random  
5-500 Hz: 0.5 g (0 to peak), swept sine  
5-500 Hz: 2.09 g RMS, random  
125 G, 8.6 m/s, trapezoidal pulse  
1.6 m/s, <3 msec duration, half-sine pulse  
30 minutes  
Samples of this product have been type tested in accordance with the Keysight environmental test manual and verified to be robust against the  
environmental stresses of storage, transportation and end-use; those stresses include but are not limited to temperature, humidity, shock, vibration,  
altitude, and power line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class3.  
Regulatory  
Safety  
IEC/ EN 61010-1  
USA: ANSI/UL 61010  
EMC  
IEC 61326-1  
Module Calibration  
Both adjustment and performance verification is required to calibrate M9195A modules. Self-maintainers can perform these proce-  
dures by using software available from Keysight:  
– N7800A Test management environment  
– N7876A PXI digital IO calibration application  
Module adjustment requires the Y1252A adjustment cable. This accessory consists of three cable assemblies necessary for voltage and  
current adjustments of the M9195A. These cables are used with a high-end digital multimeter (DMM) to adjust on-board analog-digital  
conversion. The N7876A is required to make adjustments using this cable.  
Other instruments and hardware is required for calibration. The N7876A WebHelp file lists the equipment required for adjustment and  
verification of the M9195A when using the N7800A and N7876A. For more details and access to the N7876A WebHelp file see:  
http://cal.software.keysight.com/.  
 复制成功!