Technical specifications and characteristics  
PPMU Characteristics  
Characteristic  
Value  
Comments  
* denotes warranted specification  
Number of PPMU channels  
PPMU modes  
16  
Force V measure V, Force V measure I, Force I mea-  
sure I, Force I measure V, Force nothing measure V  
Measurement averaging modes  
Force voltage range  
None, 64 averages, 50 Hz one PLC, 60 Hz one PLC  
-2 V to +6.5 V (| Current | ≤ 4 mA)  
-2 V to +6 V (| Current | ≤ 25 mA)  
-2 V to +5.75 V (| Current | ≤ 40 mA)  
(16-bit with 152 µV resolution)  
Force voltage accuracy *  
± 10mV  
Maximum accuracy from ±5 ° C of AutoCorrections.  
Ground sense tied to ground.  
Measure voltage accuracy *  
± 10mV  
With remote sense. Maximum accuracy from ±5 ° C  
of AutoCorrections with one PLC averaging at sense  
location. Ground sense tied to ground  
Force voltage settling time  
< 20 µS (40 mA range)  
< 20 µS (1 mA range)  
< 25 µS (100 µA range)  
< 100 µS (10 µA range)  
< 525 µS (2 µA range)  
1 V rising and falling step settled to 1% of final value  
into 1M Ω/ 1nF load. Typical  
Force voltage stability  
Current range  
Stable at all ranges into 1 µF  
Larger load capacitance possible, but response  
limited by current slew rate.  
-40 mA to + 40 mA  
(16-bit with 2.44 µA resolution)  
Maximum of 250 mA per module in combination  
with other channels  
-1 mA to + 1 mA  
(16-bit with 61 nA resolution)  
-100 µA to + 100 µA  
(16-bit with 6.1 nA resolution)  
-10 µA to + 10 µA  
(16-bit with 610 pA resolution)  
-2 µA to + 2 µA  
(16-bit with 122 pA resolution)  
Force current accuracy *  
± 1% of range  
± 1% of range  
Maximum accuracy from ±5 ° C of AutoCorrections.  
Remote sense disconnected.  
Measure current accuracy *  
Maximum accuracy from ±5 ° C of AutoCorrections  
with one PLC averaging. Remote sense disconnect-  
ed.  
Force current settling time  
Channel leakage  
Dependent on load  
± 10 nA  
Typical  
Remote sense  
16 channels  
1 GND  
Ground sense should be tied to ground at  
measurement location for maximum accuracy.  
Remote sense leakage  
± 4 nA  
Typical  
Force current voltage clamp range  
-2 V to +6.5 V  
VCH > VCL  
(16-bit with 152 µV resolution)  
VCL (-2 V to +4 V)  
VCH (0 V to +6.5 V)  
Force current voltage clamp accuracy  
± 50 mV  
Characteristic accuracy from ±5 ° C of  
AutoCorrections  