07 Keysight M9186A PXI Voltage/Current Source Data Sheet
Definitions for specifications
Nominal (nom):
Specification (spec):
The expected mean or average performance, or an
attribute whose performance is by design, such as the 50
Ω connector. This data is not warranted and is measured at
room temperature (approximately 25 °C).
Represents warranted performance of a calibrated
instrument that has been stored for a minimum of two
hours within the operating temperature range of 0 to
40 °C, unless otherwise stated, and after a 45-minute
warm-up period. The specifications include measurement
uncertainty. Data represented in this document are
specifications unless otherwise noted.
Measured (meas):
An attribute measured during the design phase for
purposes of communicating expected performance, such
as amplitude drift vs. time. This data is not warranted and
is measured at room temperature (approximately 25 °C).
Typical (typ):
Represents characteristic performance,
which 80% of the instruments manufactured will meet.
This data is not warranted, does not include measurement
uncertainty, and is valid only at room temperature
(approximately 25 °C).
Note:
All graphs contain measured data from several units at
room temperature unless otherwise noted.