06 | Keysight | M9111A PXIe High-Speed Source/Measure Unit - Data Sheet
Supplemental Characteristics
M9111A
Minimum current and voltage compliance limits:
13 V and 6 V ranges current priority mode
3 A range voltage priority mode
1 A range voltage priority mode
Programming range and resolution:
Voltage, 13 V range
20 mV
± 20 mA
± 10 mA
0 to 13.26 V; 260 μV
0 to 6.12 V; 260 μV
Voltage, 6 V range
Current
–3.06 A to 3.06 A; 100 μA
Programming accuracy temperature coefficient per °C:
Voltage, 13 V and 6 V ranges
Current
0.0025% +200 µV
0.0025% +60 µA
Measurement resolution:
Voltage
120 μV
25 μA
13 nA
1.4 nA
Current, 3 A range
Current, 1 mA range
Current, 100 μA range
Measurement accuracy temperature coefficient per °C:
Voltage
0.003 % + 75 µV
Current, 3 A range
Current, 1 mA range
Current, 100 uA range
0.002% + 110µA
0.002% + 5 nA
0.002% + 500 pA
Voltage programming speed and settling time: (at the specified bandwidth)
With slew rate set to maximum; with high 2 output capacitor = 1 μF; with high 3 output capacitor = 7 μF
Compensation setting
Low
High 1
High 2
High 3
Rise time from 10% to 90% of step
13 V and 6 V ranges with a 0-4 V step
300 μs
12 μs
15 μs
40 μs
Settling time to 0.1% of step
1.2 ms
40 μs
50 μs
120 μs
13 V and 6 V ranges with a 0-4 V step
High frequency output voltage noise: (at the specified bandwidth)
With high 2 setting, output capacitor = 1 μF; with high 3 setting, output capacitor = 7 μF
Compensation setting
Low
High 1
15 mV
High 2
3 mV
High 3
3 mV
CV peak-to-peak from 20Hz - 20MHz
13 V and 6V ranges with any load
15 mV
CV rms from 20Hz – 20MHz
1.5 mV
1.5 mV
0.4 mV
0.3 mV
13 V and 6 V ranges with any load
Voltage measurement noise: (peak value)
13 V range
2.9 mV
Voltage measurement settling time: (settling band of 0.1% of step - 10mV)
Settling time
35 μs
Voltage measurement small signal bandwidth:
–3 db
–1 db
DC to 30 kHz
DC to 17 kHz