05 | Keysight | M9111A PXIe High-Speed Source/Measure Unit - Data Sheet
Performance Specifications
Unless otherwise noted, specifications warranted over the ambient temperature range of 0 to 45 °C after a 30-minute warm-up period.
M9111A
DC ratings:
Voltage
Current
Power
13 V/6 V
± 1 A/± 3 A
13 W/18 W
Output voltage ripple and noise (PARD) from 20 Hz – 20 MHz:
Measured at the output terminals, under all load conditions, in voltage priority mode with output bandwidth setting = low
CV peak-to-peak
CV rms
15 mV
1.5 mV
Load effect (load regulation):
For any load change, with a lead drop of up to 1.0 V. The load lead drop reduces the maximum available voltage at the load.
Voltage
500 μV
200 μA
250 nA
Current, 3 A range
Current, 1 mA range
Programming accuracy @ 23 °C ± 5 °C:
Applies from minimum to maximum programming range at any load.
Voltage, 13 V and 6 V ranges
Current, 3 A range
0.025% + 1 mV
0.05% + 1 mA
0.05% + 500 nA
Current, 1 mA range
Measurement accuracy @ 23 °C ± 5 °C:
Applies when measuring the default value of 3255 data points with a 5.12 μs time interval.
Voltage, 13 V and 6 V ranges
Current, 3 A range
0.05% + 1 mV
0.05% + 300 μA
0.05% + 100 nA
0.05% + 10 nA
Current, 1 mA range
Current, 100 μA range
Load transient response time in voltage priority mode:
Time to recover to within the settling band.
With 150 μF cap (ESR = 50 mΩ) at load, remote sensing at cap, 4.25’ twisted pair load leads.
Rise time (10% to 90%)
≥ 10 μs
Settling band
6 V range with a 1.4 A load step
± 20 mV
Recovery time
≤ 35 μs