08 | Keysight | M9111A PXIe High-Speed Source/Measure Unit - Data Sheet
Supplemental Characteristics (continued)
M9111A
Current measurement small signal bandwidth:
–3 dB
3 A range
1 mA range
100 μA range
DC to 28 kHz
DC to 10 kHz
DC to 1 kHz
–1 dB
3 A range
100 mA range
100 μA range
DC to 16 kHz
DC to 6 kHz
DC to 800 Hz
Current priority transient characteristic: (with 4.25' twisted pair load leads)
3A range with a 1.5-3.5 V step
Current settling band
Recovery time
40 mA
5 μs
Common mode current from 20 Hz – 20 MHz: (with negative output connected to chassis)
CC peak-to-peak
CC rms
< 1 mA
< 75 μA
Remote sense capability:
Outputs can maintain specifications with up to a 1-volt drop per load lead. The load lead drop reduces the maximum available voltage at the load.
Measurement digitizer:
Digitize voltage and current at ~200 ksamples/s, 1M readings
Regulatory compliance:
EMC
Complies with European EMC Directive 2014/30/EU
– IEC/EN 61326-1
– CISPR 11, Group 1, class A
– AS/NZS CISPR 11
Complies with South Korea KC mark
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Safety
Complies with European Low Voltage Directive 2014/35/EU.
Environmental conditions:
Temperature range
Relative humidity
Altitude
0°C to 45°C
Up to 95% (non-condensing)
Up to 2000 meters
–40°C to 70°C
Storage temperature
Output terminal isolation:
Maximum rating
Net weight:
No output terminal may be more than 60 VDC from any other terminal or chassis ground.
0.41 kg (0.90 lbs)
Dimensions:
3U, 1-slot, PXIe module
19.9 mm W x 128.4 mm H x 212.6 mm D (0.784 in. x 5.06 in. x 8.37 in.)