05 | Keysight | M9111A PXIe High-Speed Source/Measure Unit - Data Sheet  
Performance Specifications  
Unless otherwise noted, specifications warranted over the ambient temperature range of 0 to 45 °C after a 30-minute warm-up period.  
M9111A  
DC ratings:  
Voltage  
Current  
Power  
13 V/6 V  
± 1 A/± 3 A  
13 W/18 W  
Output voltage ripple and noise (PARD) from 20 Hz – 20 MHz:  
Measured at the output terminals, under all load conditions, in voltage priority mode with output bandwidth setting = low  
CV peak-to-peak  
CV rms  
15 mV  
1.5 mV  
Load effect (load regulation):  
For any load change, with a lead drop of up to 1.0 V. The load lead drop reduces the maximum available voltage at the load.  
Voltage  
500 μV  
200 μA  
250 nA  
Current, 3 A range  
Current, 1 mA range  
Programming accuracy @ 23 °C ± 5 °C:  
Applies from minimum to maximum programming range at any load.  
Voltage, 13 V and 6 V ranges  
Current, 3 A range  
0.025% + 1 mV  
0.05% + 1 mA  
0.05% + 500 nA  
Current, 1 mA range  
Measurement accuracy @ 23 °C ± 5 °C:  
Applies when measuring the default value of 3255 data points with a 5.12 μs time interval.  
Voltage, 13 V and 6 V ranges  
Current, 3 A range  
0.05% + 1 mV  
0.05% + 300 μA  
0.05% + 100 nA  
0.05% + 10 nA  
Current, 1 mA range  
Current, 100 μA range  
Load transient response time in voltage priority mode:  
Time to recover to within the settling band.  
With 150 μF cap (ESR = 50 mΩ) at load, remote sensing at cap, 4.25’ twisted pair load leads.  
Rise time (10% to 90%)  
≥ 10 μs  
Settling band  
6 V range with a 1.4 A load step  
± 20 mV  
Recovery time  
≤ 35 μs