07 | Keysight | PXI Multiplexer Switch Modules M9101A, M9102A, M9103A – Data Sheet  
Specifications and Characteristics  
M9103A PXI high-density multiplexer  
M9103A specifications and characteristics  
The M9103A is a high-density, 99-channel, high-powered  
armature relay multiplexer. Each channel has a high and low for  
2-wire switching and can switch up to 100 Vrms, with up to 60 W  
of power. The modules’ isolation switch connects the closed  
channels to the module common. Easily connect to the multiplexer  
with a high-density, 200-pin low force helix (LFH) connector or  
cable.  
General specifications  
Channels  
99  
Switch type  
Max volts1  
2-wire armature  
100 Vrms  
1 A  
Max current switch/carry  
Switching characteristics, nominal  
Max power  
60 W  
Switch speed (typical)  
4.5 msec  
< 500 mΩ  
Initial path resistance,  
differential (typical)2  
Connectors  
Bandwidth  
200 LFH connector block or cable  
1 MHz  
DC isolation, Ch-Ch, Ch-Gnd  
25C/40%RH (typical)  
25C/80%RH (typical)  
40C/80%RH (typical)  
1x109 Ω  
1x107 Ω  
1x105 Ω  
Thermal offset, differential  
5 µV  
(typical)  
Relay life, operations  
Low power load (typical)  
Rated power load (typical)  
> 2.5x106  
> 1x105  
Figure 5. M9103A 99-ch, 2-wire, EM relays  
1. Not for connection to mains.  
2. Channel resistance is typically < 1.5 Ω but can go as high as 50 Ω when  
a channel is used in measurement applications with < 10 mA load  
current. Increased relay channel resistance for measurements with load  
currents below 10 mA can occur on cards that have been out of service  
or following relay inactivity for periods greater than 1 week. Switching  
relays for 2K cycles prior to use may reduce the variation in channel  
resistance. Keysight recommends the use of 4-wire Ohms for resistance  
measurements. For high accuracy voltage measurements, select the  
DMM input resistance setting of >10 G ohms to minimize the impact of  
relay contact resistance.  
 复制成功!