15 | Keysight | M8195A 65 GSa/s Arbitrary Waveform Generator and M8197A Multi-Channel Synchronization Module - Data Sheet  
Trigger/gate input  
A trigger/gate input is provided on the front panel of the M8195A and the M8197A. The trigger input is used to start/gate waveform  
playback.  
– The trigger/gate input of the M8195A affects the channels of that M8195A.  
– The trigger/gate input of the M8197A affects all channels of all M8195A that are combined in a multi-module system. To achieve  
best delay accuracy between trigger/gate input and the OUT signals, use the trigger/gate input of the M8197A. See timing  
characteristics below.  
M8195A and M8197A  
Input range  
-4 V to +4 V  
Threshold  
Range  
-4 V to +4 V  
Resolution  
10 mV (nom)  
Sensitivity  
100 mV (typ)  
Polarity  
Selectable, positive or negative  
50 (nom), DC coupled  
Input impedance  
Max. toggle rate1  
Sample rate divider = 1  
Sample rate divider = 2  
Sample rate divider = 4  
Connector type  
Sample rate/3584  
Sample rate/1792  
Sample rate/896  
SMA (female)  
Event input  
An event input is provided on the front panel of the M8195A and the M8197A. It is used in conjunction with the sequencer to advance  
to the next waveform segment or sequence.  
– The event input of the M8195A affects the channels of that M8195A.  
– The event input of the M8197A affects all channels of all M8195A that are combined in a multi-module system.  
M8195A and M8197A  
Input range  
Threshold  
-4 V to +4 V  
Range  
-4 V to +4 V  
Resolution  
Sensitivity  
Polarity  
10 mV (nom)  
100 mV (typ)  
Selectable, positive or negative  
50 Ω (nom), DC coupled  
SMA (female)  
Input impedance  
Connector type  
1. Without option -FSW the trigger interval is limited as described under the ´Frequency switiching characteristics´  