20 | Keysight | InfiniiVision 1000 X-Series Oscilloscopes - Data Sheet
Performance Characteristics (Continued)
Trigger system
EDUX1002A/EDUX1002G/DSOX1102A/DSOX1102G
Trigger sources
Trigger modes
Analog channel (1, 2), line 5, external, WaveGen, WaveGen modulation FM/FSK
Normal (triggered): Requires trigger event for oscilloscope to trigger
Auto: Triggers automatically in absence of trigger event
Single: Triggers only once on a trigger event
Force: Front panel button that forces a trigger
DC: DC coupled trigger
Trigger coupling
AC: AC coupled trigger, cutoff frequency: ~ 10 Hz
HF reject: High frequency reject, cutoff frequency ~ 50 kHz
LF reject: Low frequency reject, cutoff frequency ~ 50 kHz
Noise reject: Selectable OFF or ON, decreases sensitivity 2x
60 ns to 10 s
Trigger holdoff range
Trigger sensitivity
EDUX1002A/EDUX1002G
Greater of:
DSOX1102A/DSOX1102G
Greater of:
Internal 6
0.6 div or 2.5 mV (≤ 10 MHz)
0.9 div or 3.8 mV (10 to 70 MHz)
1.2 div or 5 mV (70 to 100 MHz)
0.6 div or 2.5 mV (≤ 10 MHz)
0.9 div or 3.8 mV (10 to 70 MHz)
1.2 div or 5 mV (70 to 100 MHz)
External
≤ 10 MHz:
250 mVpp
≤ 10 MHz:
50 mVpp (1.6 V range)
250 mVpp (8 V range)
100 mVpp (1.6 V range)
500 mVpp (8 V range)
10 to 50 MHz:
500 mVpp
10 to 100 MHz:
Trigger level range
EDUX1002A/EDUX1002G
± 6 div from center screen
± 8 V
DSOX1102A/DSOX1102G
± 6 div from center screen
± 1.6 V or ± 8 V selectable
Internal
External 7
5. Line trigger to ≤ 60 Hz.
6� Denotes warranted specifications; All others are typical� Specifications are valid after a 30-minute warm-up period and from ± 10 °C firmware calibration
temperature�
7� Input voltage must remain within these limits for proper operation�