23 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Table 48. Option E5092A-020 port performance (continued)
Description
Specification
Typical
Insertion loss
SPDT switch1
50 MHz to 100 MHz
100 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 14 GHz
14 GHz to 18 GHz
18 GHz to 20 GHz
4 dB
3.5 dB
4.5 dB
5 dB
5.5 dB
7 dB
8 dB
8.5 dB
10 dB
11.5 dB
SP4T switch2
50 to 100 MHz
100 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 14 GHz
14 GHz to 18 GHz
18 GHz to 20 GHz
4 dB
3.5 dB
4.5 dB
5.5 dB
6 dB
7.5 dB
8.5 dB
9.5 dB
10.5 dB
12 dB
Stability per switch
Condition: Enviroment Temperature +23 °C ± 3 °C and Internal DC source: ≤ 100 mA (Sum of 4 channels), no heat source and no wall close to the unit.
50 M to 4 GHz
4 G to 12 GHz
12 G to 20 GHz
Condition: besides the above
50 M to 4 GHz
0.003 dB/°C (SPD)
0.005 dB/°C (SPD)
0.008 dB/°C (SPD)
0.007 dB/°C (SPD)
0.012 dB/°C (SPD)
0.017 dB/°C (SPD)
4 G to 12 GHz
12 G to 20 GHz
Isolation3
50 MHz to 500 MHz
500 MHz to 1 GHz
1 GHz to 2 GHz
2 GHz to 6 GHz
6 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
65 dB
80 dB
85 dB
90 dB
85 dB
75 dB
65 dB (Over arbitrarily test ports)
1. SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.)
2. SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.)
3. This specification is defined when all ports are terminated with a 50 ohm load.