19 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Dimensions (rear view with option 1E5, in millimeters)
Measurement Throughput Summary
Measurement throughput data is supplemental performance data.
Common condition for the measurement throughput data:
– Analyzer display turned off with: DISPlay:ENABle OFF, DISPlay:VISible OFF
– Number of traces = 1
– firmware version: A.11.00
Cycle time for measurement completion
Sweep type: Auto
System error correction: OFF
Sweep type: Stepped
System error correction: ON
Number of Points
51
201
401
1601
51
201
401
1601
Start 800 MHz, stop 1 GHz, 500 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
1.3
2.2
4.0
1.8
3.0
6.2
2.5
5.2
12
7.0
17
31
1.5
2.6
4.7
2.8
5.1
11
4.3
8.2
18
13
24
51
Start 800 MHz, stop 1 GHz, 1 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
51
196
391
781
389
777
1546
3092
6187
51
196
391
781
389
777
1546
3092
6187
101
201
101
201
1554
1554
Start 100 kHz, stop 4.5 GHz, 500 kHz IF bandwidth
Uncorrected
2-port cal
4-port cal
3.1
5.7
11
5.4
11
7.7
15
31
12
24
50
3.1
5.8
11
5.4
11
7.8
15
31
20
39
81
21
21
Unit: ms