12 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Dynamic accuracy1, 2
Description
Specification
Typical
Magnitude
10 dBm
5 dBm
0 dBm
—5 dBm
—10 dBm (Ref.)
—15 dBm
—20 dBm
—30 dBm
—40 dBm
—50 dBm
—60 dBm
—70 dBm
—80 dBm
—90 dBm
—100 dBm
± 0.207 dB
± 0.071 dB
± 0.027 dB
± 0.012 dB
± 0.006 dB
± 0.009 dB
± 0.011 dB
± 0.014 dB
± 0.018 dB
± 0.021 dB
± 0.029 dB
± 0.072 dB
± 0.116 dB
± 0.159 dB
± 0.202 dB
Phase
10 dBm
5 dBm
0 dBm
—5 dBm
± 5.02 deg
± 0.67 deg
± 0.14 deg
± 0.06 deg
± 0.04 deg
± 0.05 deg
± 0.06 deg
± 0.08 deg
± 0.10 deg
± 0.13 deg
± 0.18 deg
± 0.47 deg
± 0.76 deg
± 1.05 deg
± 1.34 deg
—10 dBm (Ref.)
—15 dBm
—20 dBm
—30 dBm
—40 dBm
—50 dBm
—60 dBm
—70 dBm
—80 dBm
—90 dBm
—100 dBm
1. Accuracy of the test port input power reading is relative to –10 dBm reference input power level. IFBW is equal to or lower than 10 Hz.
2. Applies to the units with Serial Number Prefix MY552/SG552 and above