10 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet
Test Port Input
Description
Specification
Typical
Damage Level
+27 dBm or ± 35 VDC
Test port noise floor1
9 k to 100 kHz
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
—120 dBm
—127 dBm
—130 dBm
—130 dBm
—122 dBm
—130 dBm
—137 dBm
—140 dBm
—136 dBm
—128 dBm
Compression level
(@+10dBm input)
Magnitude
9 k to 6 GHz
6 G to 9 GHz
Phase
0.207 dB
0.207 dB
0.04 dB
0.08 dB
9 k to 6 GHz
6 G to 9 GHz
0.503 deg
0.503 deg
0.3 deg
0.6 deg
0.1 dB compression input level
9 k to 100 k
100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
10 dBm
12 dBm
15 dBm
12 dBm
10 dBm
1. IFBW=1 Hz, FOM is off.